Skip to main content
. 2019 Aug 27;12(17):2749. doi: 10.3390/ma12172749

Table 3.

The Max EQS and minimum principal strain (compressive strain) results (mean ± SEM) of IS-II and IS-II fixtures for cortical bone, cancellous bone, and the nerve surface (* p-value < 0.01).

Components Direction Maximum Equivalent Stress (MPa)
IS-II n IS-III n p -value
Cortical bone 100 N (Axial) 9.56 ± 0.10 4586 8.78 ± 0.09 3906 <0.001 *
100 N (15°) 22.07 ± 0.32 19.70 ± 0.29 <0.001 *
30 N (45°) 9.28 ± 0.12 8.38 ± 0.10 <0.001 *
Cancellous bone 100 N (Axial) 0.24 ± 0.002 8192 0.25 ± 0.002 7450 <0.001 *
100 N (15°) 0.72 ± 0.005 0.76 ± 0.005 <0.001 *
30 N (45°) 0.29 ± 0.002 0.31 ± 0.002 <0.001 *
Components Direction Maximum Equivalent Stress (kPa)
IS-II n IS-III n p -value
Nerve 100 N (Axial) 18.80 ± 0.10 6031 18.30 ± 0.09 6031 <0.001 *
100 N (15°) 19.80 ± 0.21 19.40 ± 0.20 0.168
30 N (45°) 8.90 ± 0.07 8.80 ± 0.07 0.775
Components Direction Minimum Principal Strain (×10−3)
IS-II n IS-III n p- value
Nerve 100 N (Axial) 2.796 ± 0.015 6031 2.719 ± 0.015 6031 <0.001 *
100 N (15°) 2.339 ± 0.025 2.278 ± 0.025 0.088
30 N (45°) 0.759 ± 0.007 0.751 ± 0.006 0.408