Table 1. Sensitivity Comparison for Constant Sample Sizea.
| S-band |
X-band |
|||
|---|---|---|---|---|
| mode of detection | freq (MHz) | S/N | freq (MHz) | S/N |
| regular EPR with field modulation | 4037 | 978 | 9871 | 1704 |
| regular cavity with VST | 4038 | 714 | 9871 | 1250 |
| wire micro strip resonator with VST | 3683 | 237 | 9829 | 321 |
The signal of a dpph powder sample of 4 μL volume was measured under standardized conditions of 100 G scan range, −20 dBm microwave power, no RC-filtering, 120 s acquisition time, and (when applicable) 0.5 G modulation, with the indicated setups. All resulting 1024-point spectra were subjected to 6-point Savitzky–Golay filtering. Signal-to-noise (S/N) values are reported as ratios of the peak-to-peak amplitudes of the signal and the noise in the first-derivative spectrum.