Skip to main content
. 1981 Sep-Oct;86(5):457–493. doi: 10.6028/jres.086.020

Table 1.

Calibration Constants of Wires.

Temperature Range of L.S. Fit Number of Points in Fit Applicable Temperature Range A(Ω) B(Ω K−1) Coefficients C(Ω K−2) D(Ω (MPa)−1) Standard Deviation of Fit, One Sigma Equivalent Error in T
Long wire 76 – 158 K 707 77 – 150 K −10.231206 0.3670809 −0.9903371 × 10−4 −7.796325 × 10−4 0.01 Ω 0.023 K
142 – 305 K 1684 150 – 315 K − 9.065472 0.3534445 −0.5923443 × 10−4 −1.401463 × 10−3 0.019 Ω 0.054 K
Short wire 76 – 158 K 707 77 – 150 K − 5.057558 0.1823527 −0.5257280 × 10−4 −4.144377 × 10−4 0.01 Ω 0.059 K
142 – 305 K 1684 150 – 315 K − 4.346459 0.1740251 −0.2831553 × 10−4 −6.565822 × 10−4 0.011 Ω 0.065 K