Table 1.
Measurement No. | Nominal source activity (mCi) | Nominal chamber-to-source distance (meters) | Measured multiplied by D2 (mR · m2 · h−1) |
A verage (mR · m2 · h−1) |
Avg. corrected for room scatter (mR · m2 · h−1) |
---|---|---|---|---|---|
1 | 500 | 0.5 | 137.94 | ||
2 | 500 | .5 | 137.62 | 137.83 | 137.20 |
3 | 500 | .5 | 137.92 | ||
4 | 500 | .5 | 137.36 | ||
5 | 500 | .5 | 136.82 | 137.14 | 136.51 |
6 | 500 | .5 | 137.24 | ||
7 | 500 | 1.0 | 139.58 | ||
8 | 500 | 1.0 | 140.03 | 139.80 | 137.27 |
10 | 50 | 0.5 | 15.05 | ||
11 | 50 | .5 | 14.89 | 14.97 | 14.90 |
12 | 5 | .5 | 1.34 | 1.33 | |
13 | 100 | .5 | 36.11 | 35.94 |