Skip to main content
. 2019 Sep 11;8(5):6. doi: 10.1167/tvst.8.5.6

Table 3.

Limits of Agreement for Volume and Average Thickness in Each Sector of the 6-mm Macular Grid Between Scans That Had Automated Delineated Lines Versus Manually Corrected Lines

Layer
Full Volume
Central Macula
Superior Inner
Temporal Inner
Inferior Inner
Nasal Inner
Superior Outer
Temporal Outer
Inferior Outer
Nasal Outer
Volume, mm3
 Total retina ±0.095 ±0.007 ±0.010 ±0.005 ±0.006 ±0.010 ±0.019 ±0.010 ±0.015 ±0.055
 RNFL ±0.074 ±0.004 ±0.008 ±0.007 ±0.010 ±0.012 ±0.019 ±0.012 ±0.016 ±0.048
 IRL ±0.110 ±0.018 ±0.014 ±0.015 ±0.017 ±0.016 ±0.043 ±0.013 ±0.023 ±0.062
 ORL ±0.041 ±0.020 ±0.013 ±0.017 ±0.017 ±0.017 ±0.039 ±0.017 ±0.018 ±0.017
Average thickness, μm
 Total retina ±6.790 ±3.903 ±2.747 ±3.729 ±5.882 ±3.843 ±1.851 ±2.816 ±10.107
 RNFL ±4.348 ±3.877 ±6.761 ±2.741 ±4.358 ±3.564 ±2.367 ±2.912 ±8.384
 IRL ±9.156 ±6.893 ±7.030 ±8.128 ±7.630 ±4.420 ±2.042 ±3.556 ±10.406
 ORL ±10.402 ±5.947 ±7.505 ±7.227 ±6.509 ±1.707 ±1.622 ±2.720 ±2.298