. 2019 Sep 19;18:1533033819877986. doi: 10.1177/1533033819877986

Table 2.

Magnetic Resonance Imaging Isocenter Displacement From the Virtual Isocenter Measured With the DQA and MagPhan.a

Gantry Angle DQA MagPhan
Ave. ΔX± 1SD (mm) Ave. ΔY± 1SD (mm) Ave. ΔZ± 1SD (mm) Ave. ΔX± 1SD (mm) Ave. ΔY± 1SD (mm) Ave. ΔZ± 1SD (mm)
0.57 ± 0.12 0.63 ± 0.12 −0.20 ± 0.0 0.57 ± 0.06 0.50 ± 0.0 −0.37 ± 0.06
270° 0.43 ± 0.12 0.50 ± 0.0 0.87 ± 0.06 0.37 ± 0.06 0.50 ± 0.0 0.53 ± 0.06
180° −0.70 ± 0.0 0.50 ± 0.0 0.70 ± 0.10 −0.80 ± 0.0 0.57 ± 0.06 0.30 ± 0.0
90° −0.57 ± 0.12 0.47 ± 0.06 −0.27 ± 0.06 −0.50 ± 0.0 0.50 ± 0.0 −0.70 ± 0.0
Ave −0.07 0.53 0.28 −0.09 0.52 −0.06
Max-Min 1.27 0.17 1.13 1.37 0.07 1.23

Abbreviations: Ave, average; DQA, daily quality assurance; Max, maximum; Min, minimum; MRI, magnetic resonance imaging.

a The standard deviation is for the 3 experimental runs, while the average and range in the bottom 2 rows are for the 4 cardinal gantry angles.