Table 1.

Summary of reconstruction techniques for 2D cine CMRI

Authors Year Mode Method Salient features Performance
Tsao et al. [59] 2003 Multi-BH k-t BLAST, k-t SENSE Model-based method exploiting data correlations to recover unacquired samples. Cartesian sampling 4-fold acceleration. Spatial res., 2.42 × 2.52 mm2 (slice thickness 10 mm). Temporal res., 26 ms
Huang et al. [60] 2005 Multi-BH k-t GRAPPA GRAPPA combined with sliding window techniques for missing data interpolation. Cartesian sampling AF = 7. Reduction factor, 5.17. Spatial res., 1.77 × 1.82 mm2 (slice thickness, 6 mm). Number of phases, 14. Reconstruction time, 4 s per frame
Lustig et al. [61] 2006 Multi-BH k-t SPARSE CS-based method exploiting spatial and temporal sparsity of data. Cartesian sampling 7-fold frame-rate acceleration. Spatial res., 2.5 × 2.5 mm2 (slice thickness, 9 mm). Temporal res., 40 ms. Reconstruction time, 1 h per 64 × 64 × 64 scene
Jung et al. [62] 2009 Multi-BH k-t FOCUSS CS method with ME-MC based on block matching. Cartesian sampling AF = 6. Spatial res., 1.25 × 1.17 mm2 (slice thickness, 5 mm). 25 cardiac phases
Pedersen et al. [63] 2009 Multi-BH k-t PCA Generalization of k-t BLAST/SENSE using PCA temporal constraint. Cartesian sampling Myocardial perfusion images acquired in a pig. 8-fold acceleration. Spatial res., 1.25 × 1.25 mm2 (slice thickness, 10 mm). 64 frames
Christodoulou et al. [64] 2010 Multi-BH PSF Partially separable function reconstruction with anatomical constraints Data of rat hearts. Spatial res., 390 μm in-plane (slice thickness, 1.5 mm). Temporal res., 15 ms
Lingala et al. [65] 2011 Multi-BH k-t SLR Low-rank structure using KLT to exploit the sparsity. Cartesian sampling Cardiac perfusion MRI data. AF = 11. Matrix size, 90 × 190
Hansen et al. [66] 2012 FB real-time Temporal multi-resolution scheme combining PI with MC based on nonrigid registration. Cartesian and golden-angle radial sampling 2-fold PI acceleration. Spatial res., 1.4–1.5 × 1.9–2 mm2 (Cartesian), 1.4–1.5 × 1.4–1.5 mm2 (golden angle radial), slice thickness, 6 mm. Temporal res., 30 ms
Asif et al. [67] 2013 Multi-BH MASTER CS with ME-MC based on motion-adaptive spatio-temporal regularization. Cartesian sampling Retrospective downsampling with reduction factor up to 10. Spatial res., 1.56 × 1.37 mm2 (slice thickness, 12 mm). 16 cardiac phases
Feng et al. [68] 2013 FB real-time

k-t SPARSE

SENSE

Combination of CS and PI for real-time imaging. Cartesian sampling 8-fold acceleration. Spatial res., 2.3 × 2.3 mm2 (slice thickness, 8 mm). Temporal res., 43.2 ms. Offline reconstruction time, 4.6 min per slice
Usman et al. [69] 2013 FB MC-CS Generalized MC in CS reconstruction. Respiratory motion self-gating by low resolution virtual 2D navigator images. Golden angle radial sampling AF = 4–6. Spatial res., 1.5–2 × 1.5–2 mm2. 20 cardiac phases. Temporal res., 30–40 ms. Reconstruction time, 2–2.5 h.
Xue et al. [70] 2013 FB real-time SPIRiT non-linear reconstruction with spatial-temporal regularization (Harr wavelet transformation) and ME-MC based on non-rigid registration. Cartesian time-interleaved sampling Scan time, 16–20 s per acquired slice. PI reduction factor of R = 4. Spatial res., 1.3–1.8 × 1.8–2.1 mm2 (slice thickness, 8 mm). 30 cardiac phases. Temporal res., 34.3 ± 9.1 ms. Inline reconstruction time (Gadgetron), 80–120 s per slice
Schmidt et al. [71] 2013 FB real-time rtCS11 Real-time CS-based reconstruction with k-t regularization. Cartesian sampling Scan time, 1 heartbeat. AF = 10.9. Spatial res., 1.7 × 1.7 mm2 (slice thickness, 6 mm). Temporal res., 30 ms. Online reconstruction
Trémoulhéac et al. [72] 2014 Multi-BH k-t RPCA L+S decomposition based on RPCA with temporal FT. Variable density Cartesian and pseudo-radial sampling AF = 8. Matrix size, 128 × 128 (90 frames). Reconstruction time, 10 min
Wang et al. [73] 2014 Multi-BH CS-based reconstruction with DL. Retrospective Cartesian undersampling AF up to 8. Matrix size, 150–256 × 256–304 (14–26 frames). Reconstruction time, 11.3–24.3 min
Vincenti et al. [74] 2014 Single-BH CS-based method with Cartesian acquisition AF = 11. 3 long-axis and 4 short-axis views. Spatial res., 1.5 × 1.5 mm2 (slice thickness, 6 mm). 24 cardiac phases. Temporal res., 30 ms. BH duration, 14 s
Royuela-del Val et al. [75] 2015 Single-BH kt-WiSE MC-CS based on GW registration with SENSE. Golden angle radial sampling AF = 16. Spatial res., 2 × 2 mm2, (slice thickness, 8 mm, 12 slices). 16 cardiac phases Temporal res., 46.4 ms. BH duration, 11.1 s
Velikina et al. [76] 2015 Multi-BH MOCCO Pre-estimated low-rank temporal signal models. Variable density Cartesian sampling AF up to 15. Spatial res., 1 × 1.7 mm2 (26 and 30 cardiac phases)
Otazo et al. [44] 2015 Multi-BH L+S reconstruction. Cartesian sampling for cardiac cine. Radial sampling for abdominal and breast DCE-MRI 8-fold acceleration. Spatial res., 1.25 × 1.25 mm2 (slice thickness, 8 mm). 24 temporal frames
Poddar and Jacob [77] 2016 FB real-time SToRM Manifold smoothness regularized reconstruction with radial sampling Scan time, 42 s per slice. Spatial res., 1.17 × 1.17 mm2, (slice thickness, 5 mm, 5 slices). Temporal res., 42 ms. Reconstruction time, 24 min (l2-SToRM) and 4.9 h (l1-SToRM)
Royuela-del Val et al. [78] 2016 Multi-BH GW-CS CS method with ME-MC based on non-rigid GW registration and Cartesian sampling AF up to 12. Spatial res., 2 × 2 mm2 (slice thickness, 8 mm). 16 cardiac phases
Miao et al. [79] 2016 Single-BH LLR + FD Locally low rank with temporal finite difference and PI using golden-angle radial sampling AF = 19–23. Spatial res., 2 × 2 mm2, (slice thickness, 8 mm, 12 SA slices). Temporal res., 40 ms (19–20 time frames). BH duration, 9–13 s
Feng et al. [80] 2016 FB XD-GRASP CS-based reconstruction of extra cardio-respiratory motion states. Continuous acquisition with golden-angle trajectory Scan time, 20 s per slice. AF = 16. Spatial res., 2 × 2 mm2, (slice thickness, 8 mm, 3 SA + 1 4CH slices). Temporal res., 45 ms. 18–26 cardiac phases and 10–16 respiratory phases
Chen et al. [81] 2016 FB real-time Parallel online reconstruction using dTV and accelerated reweighted least squares algorithm. Radial sampling Matrix size, 256 × 256 × 24. Reconstruction time, 33.1 s
Royuela-del Val et al. [82] 2017 Multi-BH JW-tTV CS-MC method using Jacobian weighted temporal TV as sparse regularization term. Cartesian sampling AF = 12. FOV = 320 × 320 mm2, (slice thickness, 8 mm). 30 cardiac phases
Mohsin et al. [83] 2017 Multi-BH PRICE Implicit inter-frame MC based on patch smoothness regularization. Cartesian sampling Scan time, two heartbeats per slice. AF = 6. Spatial res., 2.5 × 2.5 mm2. 1 slice, 20 temporal frames (16 lines per frame). Reconstruction time, 7 min
Chen et al. [84] 2017 Multi-BH L+S method. RPCA inverse problem solved by IALM. Cartesian and pseudo-radial sampling AF = 6. Spatial res., 1.25 × 1.25 mm2, (slice thickness, 10 mm). 1 slice, 30 temporal frames. Reconstruction time, 2–2.2 min
Xu et al. [85] 2017 Multi-BH G-NADM, L-NADM L+S method with NADM for nonconvex RPCA Matrix size, 256 × 256. 1 slice, 24 temporal frames. Reconstruction time, 3–3.3 min
Roohi et al. [86] 2017 Multi-BH k-t MLSD Multi-dimensional L+S decomposition method. Cartesian and radial sampling Sampling rate, 0.25. Spatial res., 1.35 × 1.05 mm2, (slice thickness, 10 mm). 25 temporal frames (66 bpm). Reconstruction time, 26.64 s per slice
Royuela-del Val et al. [87] 2017 Single-BH JW-tTV-GR Adaptation of JW-tTV to golden radial acquisition pattern. Whole-heart coverage AF = 16. 12–14 SA slices. Spatial res., 2 × 2 mm2 (slice thickness, 8 mm). 13–16 cardiac phases. Temporal res., 46.4 ms. BH duration, 10–13 s
Wang et al. [88] 2017 FB real-time PDLDTV Parallel DicL and dTV method using a primal-dual algorithm. Radial sampling Sampling rate, 70% 1st frame, 15% rest. Matrix size, 256 × 256, 24 temporal frames. Reconstruction time, 2 min
Tolouee et al. [89] 2018 Multi-BH L+S method with MC based on a deformable registration method. Cartesian sampling AF = 12. Spatial res., 1.35 × 1.05 mm2, (slice thickness, 10 mm). Temporal res., 25 ms
Li et al. [90] 2018 FB real-time k-space variant reduced-FOV reconstruction. Radial sampling Spatial res., 1.7 mm2, (slice thickness, 8 mm). Temporal res., 40 ms. Reconstruction time, 2 s per frame