. 2019 Jun 20;33(7):e22940. doi: 10.1002/jcla.22940

Table 2.

Validation results for measuring zonisamide concentrations in human serum by the LTIA method

  Zonisamide concentrations measured by LTIA (µg/mL)
LLOQ QC A QC B QC C
5 8 25 50
Intra‐day        
Mean (µg/mL) 5.23 6.70 22.60 45.20
Accuracy (%) 4.67 −15.80 −9.50 −9.50
Precision (%CV) 3.56 1.80 2.50 2.40
Inter‐day        
Mean (µg/mL) 5.33 7.20 23.60 46.20
Accuracy (%) 6.67 −10.60 −5.70 −7.60
Precision (%CV) 2.48 5.20 3.70 2.20

Abbreviations: CV, coefficient of variation; LLOQ, lower limit of quantitation; LTIA, latex‐enhanced turbidimetric immunoassay; QC, quality control.