Table 1. Chip-scanning parameters for 1 kHz fixed-target data collection with the Roadrunner II goniometer.
All chips were scanned with a horizontal scanning speed of 100 mm s−1.
| Chip name | lys08 | lys09 | lys10 | lys11 | lys12 | lys13 | lys14 | lys15 | protK3 | protK4 |
|---|---|---|---|---|---|---|---|---|---|---|
| Exposure time (µs) | 5 | 5 | 5 | 1 | 1 | 1 | 1 | 1 | 1 | 1 |
| No. of horizontal scan points | 331† | 331† | 331† | 333† | 331† | 151‡ | 163‡ | 156‡ | 310† | 156‡ |
| No. of vertical scan points | 116 | 105 | 104 | 116 | 105 | 24‡ | 116 | 105 | 114 | 105 |
| Total no. of scan points | 38396 | 34752 | 34423 | 38628 | 34754 | 3580 | 18907 | 16365 | 35340 | 16379 |
| No. of hits | 12209 | 12512 | 7489 | 5376 | 4621 | 937 | 4707 | 3443 | 2538 | 640 |
| No. of indexed and merged hits | 9238 | 8813 | 6293 | 4448 | 3885 | 762 | 3386 | 2312 | 1366 | 219 |
| Total scanning time (s) | 158 | 143 | 142 | 158 | 143 | 28 | 139 | 125 | 153 | 125 |
| Hits per second | 77 | 87 | 53 | 34 | 32 | 33 | 34 | 28 | 17 | 5.1 |
| Indexed patterns per second | 58 | 62 | 44 | 28 | 27 | 27 | 24 | 18 | 8.9 | 1.8 |
| Effective scanning rate (frames s−1) | 243 | 243 | 243 | 244 | 243 | 130 | 137 | 131 | 231 | 132 |
15 scan points at the beginning and end of every line were used for acceleration and deceleration of the linear axis, so the total horizontal scanning range slightly exceeds the chip lengths.
These chips were only partially scanned.