Skip to main content
. 1982 May-Jun;87(3):211–235. doi: 10.6028/jres.087.016

Table 3.

Absorbed dose measurement test results in the C, J, and C + J modes.

Mode of operation Thermistor C J power (μW) (μW) Dose rate (mGy/s) Mean error of the mean (%) Number of runs Total time (min) Potential range (V) Current range (mA)
C+J 9 9 24.93 0.1 9 40 0-0 0-0
J 9 9 24.83 .1 6 25 0-0 0-0
J 0 9 25.02 .4 6 25 0-0 0-0
24.93
±0.2%
Water agitated for 8 min, temperature raised
C+J 9 9 24.97 0.2 10 55 10-35 0.5-2
C 9 9 24.87 .3 6 35 30-20 1.5-1
C 9 0 25.08 .4 7 35 10-5 0.5-0.2
C+J 9 9 24.98 .2 7 30 0-0 0-0
24.95
±0.3%
C+J 158 158 24.95 0.7 10 85 35-15 2-1
J 158 158 24.90 .7 7 30 0-0 0-0
J 0 158 24.97 .4 6 32 0-0 0-0
24.93
±0.6%
J 160 9 24.80 0.2 10 55 25-50 1.5-3
J 0 9 24.90 .2 10 55 50-35 2.5-2
J 160 9 24.98 .2 10 40 0-0 0-0
24.90
±0.2%