. 1982 May-Jun;87(3):211–235. doi: 10.6028/jres.087.016

Table 7.

Uncertainties in the absorbed dose rate measurements.

Source Estimated uncertainty (%)
Statistical (degrees of freedom) si(vj) Other uj
Reproducibility of water 0.2 (8)
Sensitivity of C-thermistor 0.1 (3) 0.1
Sensitivity of J-thermistor 0.1 (3) 0.1
Effect of lead resistances 0.05
Beam attenuation of calorimeter lid 0.05 (5)
Beam exposure timing 0.05
Extrapolation to zero thermistor power 0.3
Polyethylene films 0.05

Combined uncertainty = Σsi2+Σuj2=0.4%