Skip to main content
. 1983 Mar-Apr;88(2):105–116. doi: 10.6028/jres.088.006

Table 5.

Measurements on ampoules of n-Heptane.

A) The data
Set No.—
Ampoule
% purity
(% purity-99.99)·104
2–05 99.9880 −20
3–04 99.9909 9
20–20 99.9956 56
7–07 99.9908 8
20–03 99.9901 1
14–10 99.9928 28
2–15 99.9915 15
8–14 99.9899 − 1
14–20 99.9906 6
7–17 99.9894 − 6
B) Biweight iterations: c =5, T(0)=7.0, s(0)=12.0, sbi=17.2
Columns give weights wi(k) corresponding to each observation yi at kth iteration. T(k)=wi(k)yi/wi(k)
obsv’n (1) (2) (3) (4) (5)
−20 .8120 .8082 .8075 .8074 .8074
9 .9989 .9992 .9992 .9993 .9993
56 .4546 .4597 .4606 .4607 .4608
8 .9997 .9999 .9999 .9999 .9999
1 .9903 .9893 .9891 .9891 .9891
28 .8839 .8869 .8875 .8876 .8876
15 .9827 .9839 .9841 .9842 .9842
− 1 .9827 .9815 .9812 .9812 .9812
6 .9997 .9996 .9995 .9995 .9995
− 6 .9547 .9527 .9523 .9523 .9523
T(1)=7.283 T(2)=7.334 T(3)=7.344 T(4)=7.345 T(5)=7.346 sbi=18.648
Final location and scale estimates (original scale):
μ^ = 99.9900 + 7.346/104 = 99.9907
σ^ = 18.648/104 = .0019