Table 5.
Effects of moisture and strain on IPRTs. Underlined numbers are the numbers of sets IX and X which had been previously measured in sets III and VI
Manufacturer | Dry and no strain, Sets I-X | Total number of IPRTs from each manufacturer | Number of IPRTs showing effects of moisture | Number of IPRTs showing effects of strain | Magnitude of effect (equivalent temperature changes) | |||||
---|---|---|---|---|---|---|---|---|---|---|
Moisture (mK) | Strain (mK) | |||||||||
A | 6 | 43 | 23 | 14 | 15, 17, 10, 10, 10, 10, 17, 16, 6, 10, 3, 15, 8, 5, 5, 5, 8, 8, 10, 10, 7, 4, 2 | 15, 5, 5, 5, 3, 8, 4, 3, 4, 3, 2, 1, 3, 2 | ||||
B | 0 | 19 | 17 | 10, 5, 10, 10, 7, 10, 15, 3, 6, 5, 2, 18, 12, 8, 9, 4, 4 | 2, 1 | |||||
C | 2 | 7 | 3 | 2 | 12, 0, 2 | 3, 2 | ||||
D | 1 | 14 (1 bad and 1 open) | 11 | 0 | 20, 13, 25, 1, 8, 12, 10, 3, 10, 0, 10 | |||||
E | 0 | 11 | 11 | 0 | 10, 5, 3, 13, 13, 10, 3, 1, 10, 10, 14 | |||||
Total | 9 | 94 | 65 | 18 | ||||||
Percent | 9.6% | (1 bad and 1 open) 2.1% | 69.15% | 19.15% |