Table 4.
First series gold-marker type iodine-125 seed measurements in the re-entrant ionization chamber. Data corrected to STP and reference time 1981 June 1.5.
Source number | Mean current for random seed position (pA) | Number of seed positions | Standard deviation of the mean (%) |
---|---|---|---|
1 | 18.50 | 6 | 1.0 |
2 | 18.91 | 5 | 0.3 |
3 | 18.76 | 5 | 1.2 |
4 | 19.52 | 4 | 1.9 |