Table 2.

Temperature errors due to the uncertainty in δ for the 122.1 keV γ-ray of the 57CoFe thermometer. The temperatures were calculated for δ = +0.120(2) using the values of W(0) shown. The latter were chosen to give values of T for δ = +0.120 which cover the thermometers useful temperature range (4.0 to 76 mK). These calculations were done using Bapp = 0 and Qλ = 1.

W(0) T T′ (T′−T)/T T′ (T′−T)/T
δ = +0.120 δ = +0.122 δ = +0.118
A2 = 0.141692 A2 = 0.137795 A2 = 0.145591
A4 = 0.010013 A4 = 0.010345 A4 = 0.009687
1.18882 4.00 mK 1.91 mK −52.27% 4.89 mK 22.25%
1.14490 10.00 9.50 −4.97 10.48 4.78
1.10064 16.30 15.83 −2.85 16.75 2.78
1.08108 20.00 19.52 −2.39 20.47 2.34
1.04728 30.00 29.45 −1.84 30.54 1.81
1.02990 40.00 39.34 −1.65 40.65 1.62
1.02032 50.00 49.22 −1.55 50.77 1.53
1.01460 60.00 59.10 −1.50 60.89 1.48
1.01096 70.00 68.97 −1.47 71.02 1.45
1.00938 76.00 74.89 −1.46 77.09 1.44