Table 2.
Temperature errors due to the uncertainty in δ for the 122.1 keV γ-ray of the 57CoFe thermometer. The temperatures were calculated for δ = +0.120(2) using the values of W(0) shown. The latter were chosen to give values of T for δ = +0.120 which cover the thermometers useful temperature range (4.0 to 76 mK). These calculations were done using Bapp = 0 and Qλ = 1.
| W(0) | T | T′ | (T′−T)/T | T′ | (T′−T)/T |
|---|---|---|---|---|---|
| δ = +0.120 | δ = +0.122 | δ = +0.118 | |||
| A2 = 0.141692 | A2 = 0.137795 | A2 = 0.145591 | |||
| A4 = 0.010013 | A4 = 0.010345 | A4 = 0.009687 | |||
| 1.18882 | 4.00 mK | 1.91 mK | −52.27% | 4.89 mK | 22.25% |
| 1.14490 | 10.00 | 9.50 | −4.97 | 10.48 | 4.78 |
| 1.10064 | 16.30 | 15.83 | −2.85 | 16.75 | 2.78 |
| 1.08108 | 20.00 | 19.52 | −2.39 | 20.47 | 2.34 |
| 1.04728 | 30.00 | 29.45 | −1.84 | 30.54 | 1.81 |
| 1.02990 | 40.00 | 39.34 | −1.65 | 40.65 | 1.62 |
| 1.02032 | 50.00 | 49.22 | −1.55 | 50.77 | 1.53 |
| 1.01460 | 60.00 | 59.10 | −1.50 | 60.89 | 1.48 |
| 1.01096 | 70.00 | 68.97 | −1.47 | 71.02 | 1.45 |
| 1.00938 | 76.00 | 74.89 | −1.46 | 77.09 | 1.44 |