Table 3.
Temperature differences for three different sets of values of ΔM and P/k for the 60CoCo thermometer. All the temperatures were calculated using the values of W(0) shown. The latter were chosen to give values of T for ΔM = −6.0725 mK and P/k = −2.2 μK which cover the thermometers useful temperature range (1.3-50 mK). The values used for U2A2 and UλAλ are those given in table 1 for the 1332.5 keV transition. Solid angle correction factors were Q2 = 0.9965 and Q4 = 0.9886. See text for further details.
W(0) | T | T′ | (T′−T)/T | T′ | (T′−T)/T |
---|---|---|---|---|---|
ΔM = −6.0725 mK | ΔM = −6.0653 mK | ΔM = −6.0668 mK | |||
P/k = −2.2 μK | P/k = −2.5 μK | P/k = −2.9 μK | |||
0.014932 | 1.30000 mK | 1.29904 mK | −0.074% | 1.30012 mK | +0.010% |
0.327535 | 5.00000 | 4.99606 | −0.079 | 4.99993 | −0.001 |
0.470108 | 6.90000 | 6.89437 | −0.082 | 6.89945 | −0.008 |
0.633327 | 10.00000 | 9.99147 | −0.085 | 9.99835 | −0.016 |
0.783152 | 15.00000 | 14.98679 | −0.088 | 14.99656 | −0.023 |
0.860948 | 20.00000 | 19.98236 | −0.088 | 19.99537 | −0.023 |
0.931038 | 30.00000 | 29.97466 | −0.084 | 29.99561 | −0.015 |
0.959508 | 40.00000 | 39.96856 | −0.079 | 39.99951 | −0.001 |
0.973528 | 50.00000 | 49.96406 | −0.072 | 50.00723 | +0.014 |