Skip to main content
. 1983 May-Jun;88(3):175–217. doi: 10.6028/jres.088.011

Table 4.

Temperature errors due to the uncertainties in the values of ΔM and P/k for the 60CoCo(hcp) thermometer. The same parameters were used to determine W(0) and T as those used in table 3. The quantity ΔT/T = (T′-T)/T is given in percent. See text for further details.

W(0) T ΔT/T ΔT/T
ΔM = −6.0725 mK Dipole error Quadrupole error
/k = −2.2 μK δΔM = ±2.4 μK δ(P/k) = ±0.4 μK
0.014932 1.3 mK ±0.039% ±0.059%
0.327535 5.0 ±0.039 ±0.053
0.470108 6.9 ±0.039 ±0.049
0.633327 10.0 ±0.039 ±0.044
0.783152 15.0 ±0.039 ±0.040
0.860948 20.0 ±0.039 ±0.040
0.931038 30.0 ±0.039 ±0.045
0.959508 40.0 ±0.039 ±0.053
0.973528 50.0 ±0.039 ±0.062