Table 6.
Temperature errors due to the uncertainties in the solid angle correction factors Q2 and Q4 for the 60CoCo(hcp) thermometer. The same parameters were used to determine W(0) and T as those used in table 3. See text for further details.
W(0) | T | ΔT/Ta |
---|---|---|
Q2 = 0.9965 | δQ2 = ±5 X 10−3 | |
Q4 = 0.9886 | δQ4 = ±20 X 10−3 | |
0.014932 | 1.3 | +2.063% |
−2.192 | ||
0.327535 | 5.0 | ±0.124 |
0.470108 | 6.9 | ±0.083 |
0.633327 | 10.0 | ±0.056 |
0.783152 | 15.0 | ±0.039 |
0.860948 | 20.0 | ±0.033 |
0.931038 | 30.0 | ±0.029 |
0.959508 | 40.0 | ±0.027 |
0.973528 | 50.0 | ±0.027 |
The largest errors are given, when δQ2 and δQ4 are both positive or negative.