Table 7.

Temperature errors for various values of Δθ for the 60CoCo(hcp) thermometer. The same parameters were used to determine W(0) and T as those used in table 3. See text for further details.

W(0) T ΔT/T ΔT/T ΔT/T ΔT/T
Δθ = ±0.5° Δθ = ±1.0° Δθ = ±2.0° δθ = ±5.0°
0.014932 1.3 mK 0.427% 1.689% 6.358% 30.626%
0.327535 5.0 0.026 0.103 0.413 2.589
0.470108 6.9 0.108 0.071 0.283 1.775
0.633327 10.0 0.012 0.048 0.193 1.213
0.783152 15.0 0.009 0.035 0.140 0.878
0.860948 20.0 0.007 0.030 0.119 0.749
0.931038 30.0 0.006 0.026 0.104 0.654
0.959508 40.0 0.006 0.025 0.099 0.620
0.973528 50.0 0.006 0.024 0.096 0.605