Table 9.

Temperature errors due to closure domains for the 60CoCo(hcp) thermometer. These calculations were done for the closure domain structure shown in figure 17a. The same parameters were used to calculate W(0), W(π/2), and T as those used in table 3. See text for further details.

W(0) W(π/2) T W(0)′ ΔT/T W(0)′ ΔT/T
f = 0.9995 f = 0.9975
D = 10 μ, L = 1 cm D = 50 μ, L = 1 cm
0.014932 1.249941 1.3 mK 0.015549 1.405% 0.018049 6.630%
0.327535 1.228283 5.0 0.327986 0.106 0.329787 0.531
0.470108 1.201600 6.9 0.470473 0.081 0.471936 0.406
0.633327 1.155682 10.0 0.633588 0.063 0.634633 0.314
0.783152 1.099698 15.0 0.783310 0.050 0.783943 0.253
0.860948 1.066134 20.0 0.851051 0.045 0.861461 0.227
0.931038 1.033636 50.0 0.931090 0.041 0.931295 0.206
0.959508 1.019977 40.0 0.959538 2.040 0.959653 0.199
0.97 3528 1.013122 50.0 0.973548 0.039 0.973627 0.199