Skip to main content
. 1984 Mar-Apr;89(2):143–185. doi: 10.6028/jres.089.009

Table 7.

Stability of ihe radiation intensity, expressed as photomultiplier signal average (V), determined for a copper (Cu), silver (Ag), and iron (Fe) hollow cathode operated in helium at 10 Torr and a current of 200, 100, and 200 mA. The measurements were taken at time intervals of 100 s after a preburn of 10 min.

V
Time, s Cu Ag Fe
0 8.43 8.13 6.00
100 8.43 8.15 6.05
200 8.33 8.15 5.97
300 8.20 8.17 6.00
400 8.17 8.20 5.95
500 8.23 8.20 5.90
600 8.19 8.20 5.95
700 8.09 8.19 5.87
800 8.07 8.17 5.85
900
8.10
8.27
5.85
Aver 8.22 8.18 5.94
σ 0.13 0.04 0.07
% σ 1.6 0.5 1.2