Table 3.

Summary of concordant (zero error and minor error) and nonconcordant (major error) readings of the in-house and commercial minimum inhibitory concentration strips against broth microdilution reference method

In-house strip Commercial strip

Acceptability (%) Outside±1log2 dilution (major error) Acceptability (%) Outside±1log2 dilution (major error)


Same reading as broth microdilution (zero error) Within±1log2 dilution (minor acceptable error) Same reading as broth microdilution (zero error) Within±1log2 dilution (minor acceptable error)
Standardization and optimization stage (%)
 ATCC 29213 17/30 13/30 0/30 2/30 12/30 16/30
30/30 (100) 14/30 (46.67)
 Clinical isolates 35/90 48/90 7/90 19/90 41/90 30/90
83/90 (92.2%) 60/90 (66.67%)
Validation stage (%)
 ATCC 29213 20/30 10/30 0/30 15/30 12/30 3/30
30/30 (100) 27/30 (90)
 Clinical isolates 62/90 24/90 4/90 20/90 50/90 20/90
86/90 (95.5%) 70/90 (77.78)

MIC=Minimum inhibitory concentration