ABSTRACT
Objective
A large‐scale audit and peer review of ultrasound images may improve sonographer performance, but is rarely performed consistently as it is time‐consuming and expensive. The aim of this study was to perform a large‐scale audit of routine fetal anatomy scans to assess if a full clinical audit cycle can improve clinical image‐acquisition standards.
Methods
A large‐scale, clinical, retrospective audit was conducted of ultrasound images obtained during all routine anomaly scans performed from 18 + 0 to 22 + 6 weeks' gestation at a UK hospital during 2013 (Cycle 1), to build a baseline understanding of the performance of sonographers. Targeted actions were undertaken in response to the findings with the aim of improving departmental performance. A second full‐year audit was then performed of fetal anatomy ultrasound images obtained during the following year (Cycle 2). An independent pool of experienced sonographers used an online tool to assess all scans in terms of two parameters: scan completeness (i.e. were all images archived?) and image quality using objective scoring (i.e. were images of high quality?). Both were assessed in each audit at the departmental level and at the individual sonographer level. A random sample of 10% of scans was used to assess interobserver reproducibility.
Results
In Cycle 1 of the audit, 103 501 ultrasound images from 6257 anomaly examinations performed by 22 sonographers were assessed; in Cycle 2, 153 557 images from 6406 scans performed by 25 sonographers were evaluated. The analysis was performed including the images obtained by the 20 sonographers who participated in both cycles. Departmental median scan completeness improved from 72% in the first year to 78% at the second assessment (P < 0.001); median image‐quality score for all fetal views improved from 0.83 to 0.86 (P < 0.001). The improvement was greatest for those sonographers who performed poorest in the first audit; with regards to scan completeness, the poorest performing 15% of sonographers in Cycle 1 improved by more than 30 percentage points, and with regards to image quality, the poorest performing 11% in Cycle 1 showed a more than 10% improvement. Interobserver repeatability of scan completeness and image‐quality scores across different fetal views were similar to those in the published literature.
Conclusions
A clinical audit and a set of targeted actions helped improve sonographer scan‐acquisition completeness and scan quality. Such adherence to recommended clinical acquisition standards may increase the likelihood of correct measurement and thereby fetal growth assessment, and should allow better detection of abnormalities. As such a large‐scale audit is time consuming, further advantages would be achieved if this process could be automated. © 2018 The Authors. Ultrasound in Obstetrics & Gynecology published by John Wiley & Sons Ltd on behalf of the International Society of Ultrasound in Obstetrics and Gynecology.
Keywords: 2D ultrasound, clinical audit, fetal growth, pregnancy, routine anatomy screening
INTRODUCTION
A mid‐trimester fetal anatomy ultrasound scan (or ‘anomaly scan’) is offered to pregnant women in most developed countries. Clinical guidelines, such as those defined by the Fetal Anomaly Screening Programme (FASP)1 in the UK or the International Society of Ultrasound in Obstetrics and Gynecology (ISUOG)2, specify protocols for such screening. For instance, FASP recommends ultrasound image capture of six views: the head in the transventricular (TV) and the transcerebellar (TC) planes, femur length (FL), abdominal circumference (AC), the lips in the coronal view and the spine. Additionally, FASP recommends visualization (though it does not mandate image capture) of other fetal structures, such as the heart, kidneys and limbs.
Such protocols aim to guide practice of fetal ultrasonography to provide a ‘checklist’ of what imaging planes to capture. In principle, if all standard imaging planes are examined, detection of anomalies should be maximized. The quality of archived images reflects overall quality of fetal anatomical survey. Whilst this correlation is not absolute, the concern is that failure to archive images or poor image quality could support medicolegal claims.
Regular audit, peer review and quality assurance procedures are recognized to improve and sustain good practice3, 4, 5. However, routinely auditing images is a resource‐intensive administrative task; it is time‐consuming, tedious and takes highly trained staff away from clinical work. Consequently, a comprehensive ultrasound image audit is often undertaken only in a ‘reactive’ manner to a potential ‘screening failure’, such as in response to an unanticipated major anomaly at birth. Given the relative rarity of major anomalies, such a reactive approach is unlikely to highlight poor performance of a single operator or a department. Even when departments do perform a regular image audit, this is usually performed on a limited number of scans per sonographer. Such a selective audit is incomplete and may not provide adequate insight into sonographer performance6. Finally, however desirable a comprehensive image audit is, its resource‐intensive nature means that findings and feedback may seem no longer relevant to current practice5, 7.
In theory, an automated software‐based ultrasound image audit could be a solution for full (100%) clinical auditing. Putting aside technical challenges to achieve this, it is important to first ascertain whether a comprehensive image‐quality audit can improve performance.
The aim of this study was to determine whether a full clinical audit improves performance in terms of image acquisition/storage and image quality in a large maternity ultrasound department.
METHODS
Overview of audit cycle
This quality‐improvement study comprised a retrospective baseline audit (Cycle 1) of anonymized routine fetal anomaly scans performed over a 12‐month period in 2013 at a large teaching hospital with over 6500 births per annum. This was followed by feedback to sonographers of individual and departmental performance scores and implementation of a number of targeted interventions to improve quality of imaging. In the subsequent 12‐month period, a re‐audit (Cycle 2) using the same quality criteria was performed in order to evaluate the impact of the changes.
Audit process
Ultrasound images and meta‐data (scan date, ultrasound machine used, sonographer ID) were extracted from the hospital database (ViewPoint, GE Medical Systems, Zipf, Austria) and anonymized to remove all patient identifiable information.
Images were audited by 12 experienced sonographers, currently practicing in different hospitals in the UK, using a standard audit pro forma including a custom‐built easy‐to‐use online interface created by Intelligent Ultrasound Ltd (Milton Park, Oxfordshire, UK). This facilitated the auditing of scan completeness (assessment of whether a scan had the complete minimum set of required images) and image quality (assessment of recorded images against a scoring system).
Before commencing the study, all assessors participated in a training day to ensure familiarity with imaging protocols, quality criteria and software, and had the opportunity to practice quality assessment using a large number of case examples.
Scan completeness and image quality were measured at the level of the department and the individual sonographer for all departmental sonographers who undertook at least 30 anomaly scans in each cycle.
Scan completeness
According to the UK FASP guidelines1, a complete fetal anatomy examination should include images of the following six views: head in the TV plane, head in the TC plane, AC, FL, spine and face (lips) in the coronal plane. For the purposes of this audit, if at least one required image was missing from an examination, it was considered incomplete. Scan completeness was defined as the proportion of the total number of audited patient scans that were complete (i.e. all required images were obtained). View completeness was defined as the proportion of scans that had at least one image for a particular view.
Image‐quality scoring
The quality of images of the head TV, head TC, AC, FL, face (lips) and spine views was also audited. The quality of each image was assessed against a number of criteria (Table 1) based on studies and guidelines, and reflective of the protocol1, 8, 9. A score for each image was computed as the sum of satisfied criteria in a view; the maximum achievable score was 7 for the head TV, head TC and spine images, 6 for the AC images and 4 for the FL and coronal face images. As the number of criteria varied according to view, each score was divided by the maximum score, to give a normalized score between 0 and 1. This allowed computation of an average image‐quality score overall and for each view. The median image‐quality score for each sonographer was also computed.
Table 1.
Criteria for quality assessment of ultrasound images of head transventricular (TV), head transcerebellar (TC), abdomen, femur, spine and coronal face views
| Head TV | Head TC | Abdomen | Femur | Spine | Coronal face |
|---|---|---|---|---|---|
| Symmetrical plane with midline echo of falx dividing skull | Symmetrical plane with midline echo of falx dividing skull | Stomach bubble visible | Both ends of ossified diaphysis clear | Magnification at least 30% of screen | Upper lip visible |
| Cavum septum visible one‐third of way along midline falx | Cavum septum visible one‐third of way along midline falx | Umbilical vein one‐third of way along anteroposterior diameter, at level of portal sinus | Angle of insonation 45–90° | Continuity intact and posterior skin edge visible | Two nostrils visible |
| Cerebellum not visible | Cerebellum visible | Circular plane | Magnification at least 30% of screen | Alignment of vertebrae visible | Two lip angles visible |
| Posterior ventricle visible at level of atrium | Cisterna magna visible | Kidneys not visible | Calipers placed on clear ends of diaphysis | Amniotic fluid visible beyond skin | Adequate magnification |
| Magnification at least 30% of screen | Magnification at least 30% of screen | Magnification at least 30% of screen | Lower (sacrum) visible | ||
| Head circumference calipers placed appropriately on outer parts of skull | Calipers placed correctly on outer limits of cerebellar hemispheres | Abdominal circumference calipers placed appropriately on outer parts of abdomen | Middle (thoracic/lumbar) visible | ||
| Calipers measuring lateral ventricles at level of atrium | Calipers placed correctly on outer limits of cisterna magna | Upper (cervical/thoracic) visible |
Cycle 1: feedback and quality‐improvement measures
Findings from the first cycle of the audit relating to performance of the department as a whole were fed back to the sonography team. An individualized written report was also generated for each of the sonographers, who were given the opportunity to discuss any concerns or training needs. Following this, a number of actions and interventions were agreed and undertaken, including the production and dissemination of updated and formalized anomaly scan departmental protocol, the production for each scan room of aide memoire posters of the FASP base image menu (six views) required for storage, the introduction of guidance on image storage as well as a cardiac scanning protocol with specific hands‐on training sessions, and providing supported practice for individuals with specific training needs identified in Cycle 1 of the audit.
To assess the effect of quality‐improvement interventions on the performance of sonographers in the department, all anomaly scans acquired in 2014 were audited (follow‐up audit) following the same procedures used to analyze the 2013 data.
Measuring interobserver agreement
In order to measure agreement, 10% of the audited baseline and follow‐up images were selected by computer randomization and assessed by two independent reviewers. Agreement in scan completeness was measured as the percentage of agreed images for each view (i.e. if Reviewer 1 and Reviewer 2 agreed on the view type) and a kappa statistic was calculated. The ultrasound views for which reviewers had the highest disagreement were investigated. Variability in image‐quality assessment was evaluated by comparing overall image scores. The variability for each criterion in all views was assessed to investigate which criteria contributed most to the overall variability of reviewers.
Statistical analysis
The chi‐squared test was used to assess whether changes in scan completeness between the two cycles were statistically significant. For image‐quality scoring, paired comparisons of the normalized score (from 0 to 1) were performed using the Wilcoxon matched‐pairs signed‐ranks test, as normality tests suggested evidence of a non‐normal distribution. Two‐sided t‐tests are reported and the significance level was set at < 0.05. Analysis was performed using Stata 15 (StataCorp LLC, College Station, TX, USA).
RESULTS
In Cycle 1, 103 501 images from 6257 anomaly examinations performed by 22 sonographers were assessed. Sonographers had varying working patterns (full time, part time or locums). The median number of scans per sonographer was 237 and the median number of audited images per scan was 16. In Cycle 2, 153 557 images were evaluated from 6406 anomaly examinations performed by 25 sonographers, of whom 20 also performed examinations in Cycle 1. The median number of scans per sonographer in Cycle 2 was 189 and the median number of audited images per scan was 24. All scans were performed between 18 + 0 and 22 + 6 weeks of gestation (Figure 1). Scans were acquired using a GE Voluson E6, GE Voluson E8, GE Voluson 730 (GE Medical Systems) and Hitachi Aloka ProSound Alpha 10 (Twinsburg, OH, USA) machines, none of which had automatic caliper placement.
Figure 1.

Frequency of fetal anatomical ultrasound examinations performed in a large maternity ultrasound department in 2013 (
) and 2014 (
), according to gestational age.
Scan‐completeness audit
Overall scan completeness in the baseline audit was 72%, which improved to 78% in the follow‐up audit (χ2 = 25.99, P < 0.001) with improved completeness for all views. Table 2 summarizes the overall scan completeness and completeness per view in the baseline and follow‐up audits. Figure 2 shows scan completeness for all included sonographers (n = 20) who acquired at least 30 scans in each year. For example, sonographer 1 undertook 138 scans in Cycle 1, of which 113 were complete. The scan‐completeness audit demonstrated that the improvement was particularly noticeable for those sonographers who had the lowest completeness in the baseline audit. Sonographers whose scan completeness in Cycle 1 was less than 50% had a mean scan completeness improvement from 19.0% to 60.3% in Cycle 2. While this could be purely due to ‘regression to the mean’, this is unlikely given the consistency of improvements seen.
Table 2.
Overall and per‐view scan completeness and quality score in baseline and follow‐up audits of fetal anatomical ultrasound examinations
| View | Baseline audit | Follow‐up audit | Difference (%)* |
|---|---|---|---|
| Completeness (%) | |||
| Head TV | 89.0 | 92.8 | +3.9 |
| Head TC | 87.4 | 90.7 | +3.4 |
| Abdomen | 90.9 | 93.3 | +2.4 |
| Femur | 89.9 | 92.9 | +3.0 |
| Spine | 84.7 | 91.0 | +6.4 |
| Coronal face | 83.5 | 88.9 | +5.5 |
| Overall completeness† | 72.4 | 77.8 | +5.3 |
| Quality score | |||
| Head TV | 0.81 | 0.88 | +0.07 |
| Head TC | 0.84 | 0.90 | +0.07 |
| Abdomen | 0.86 | 0.87 | +0.01 |
| Femur | 0.85 | 0.89 | +0.04 |
| Spine | 0.68 | 0.77 | +0.09 |
| Coronal face | 0.60 | 0.68 | +0.08 |
| Median overall quality score | 0.83 | 0.86 | +0.03 |
Difference = (Follow‐up – Baseline).
Proportion of complete scans for all views.
TC, transcerebellar; TV, transventricular.
Figure 2.

Scan completeness for fetal anatomical ultrasound examinations performed by 20 sonographers, in baseline (
) and follow‐up (
) audits.
Image‐quality audit
One sonographer was excluded from the image‐quality analysis due to having insufficient images for assessment in Cycle 1. For the remaining 19 sonographers who acquired at least 30 scans in each year, the median image‐quality score (for all images in all views) improved from 0.83 (interquartile range (IQR), 0.69–0.89) in 2013 to 0.86 (IQR, 0.76–0.91) in 2014 (P < 0.001; Table 2). The median normalized image‐quality scores for these 19 sonographers demonstrate that the greatest improvement was obtained for those with the lowest baseline audit score (Figure 3). For the nine sonographers with normalized image‐quality score < 0.80 in Cycle 1, the average quality score improved from 0.70 to 0.80 in Cycle 2, while those who had good image‐quality score at baseline performed similarly in the follow‐up audit.
Figure 3.

Normalized image‐quality score for fetal anatomical ultrasound examinations performed by 19 sonographers, in baseline (
) and follow‐up (
) audits.
Interobserver agreement
The overall interobserver agreement between two reviewers for scan completeness was 95.1%. The agreement varied slightly between the different views, as shown by the head TV and spine images having the least agreement between sonographers (92.1% and 93.7%, respectively; Table 3). The overall mean interobserver agreement for assessing image quality was 82.7%. Table 4 shows the interobserver agreement for quality score for each view and each criterion.
Table 3.
Interobserver agreement of 20 sonographers for assessment of fetal anatomical ultrasound scan completeness
| View | Agreement (%) |
|---|---|
| Head TV | 92.1 |
| Head TC | 96.5 |
| Abdomen | 96.5 |
| Femur | 98.8 |
| Spine | 93.7 |
| Coronal face | 96.8 |
| Overall | 95.1 |
TC, transcerebellar; TV, transventricular.
Table 4.
Interobserver agreement in assessment of quality criteria of ultrasound images of six standard views
| View/criterion | Agreement (%) |
|---|---|
| Head TV | 87.0 |
| Ventricle calipers | 75.8 |
| Head circumference calipers | 82.0 |
| Symmetrical circular brain | 84.2 |
| Cavum septum visible | 86.1 |
| Ventricles visible | 88.8 |
| Magnification | 92.5 |
| No cerebellum | 99.3 |
| Head TC | 88.0 |
| Cavum septum visible | 79.5 |
| Calipers cisterna magna | 81.4 |
| Symmetrical circular brain | 88.4 |
| Magnification | 88.9 |
| Calipers cerebellar diameter | 89.2 |
| Cerebellum visible | 93.4 |
| Cisterna magna visible | 95.0 |
| Abdomen | 83.6 |
| Abdominal circumference calipers | 69.4 |
| Umbilical vein visible | 77.5 |
| Circular | 83.4 |
| Magnification | 86.5 |
| No kidney | 90.1 |
| Stomach visible | 94.4 |
| Femur | 82.2 |
| Magnification | 68.0 |
| Femur length calipers | 78.2 |
| Clear femoral ends | 88.1 |
| Angle of insonation | 94.3 |
| Spine | 75.4 |
| Amniotic fluid visible beyond the skin | 69.1 |
| Magnification | 73.2 |
| Continuity intact and posterior skin edge visible | 73.2 |
| Alignment of vertebrae visible | 74.5 |
| Upper (cervical/ thoracic) visible | 78.1 |
| Lower (sacrum) visible | 78.4 |
| Middle (thoracic/lumbar) visible | 81.3 |
| Coronal face | 80.0 |
| Magnification | 72.1 |
| Upper lip visible | 79.5 |
| Two lip angles visible | 81.2 |
| Two nostrils visible | 87.0 |
| Overall agreement | 82.7 |
Criteria for each view are sorted in ascending order of agreement.
TC, transcerebellar; TV, transventricular.
Audit time
The mean reviewing time per image was 24 ± 20 s, but this varied between views, as seen in the high SD. The mean time for review per scan was 8.4 ± 6.1 min for assessment of both completeness (mean, 2.0 ± 1.2 min) and grading of gradable images (6.4 ± 5.5 min). Overall, the assessment of Cycle 1 took approximately 876 h while Cycle 2 took about 897 h.
DISCUSSION
In this study, we assessed the effect of a full audit cycle involving large‐scale fetal anomaly scan audit on the performance of sonographers. In this audit, fetal scans were assessed manually by experts to evaluate scan completeness and image quality. A full‐year baseline audit was performed to establish the sonographers' baseline performance. Actions were implemented in the department in the following year and a second follow‐up audit was performed. We found that both scan completeness and image quality improved significantly between the audit cycles.
Although the improvement was by a relatively small margin as a department, Cycle 2 demonstrated large improvements for those sonographers who performed least well in the first audit cycle. Sonographers whose scan completeness in the baseline audit was less than 50% had a mean scan completeness improvement from 19.0% to 62.3% in the follow‐up audit, including two sonographers (sonographers 19 and 20) who had 0% scan completeness in the baseline audit. Similarly, normalized image‐quality score for sonographers whose baseline image‐quality score was less than 0.80 improved from a mean of 0.70 to 0.80 in the follow‐up audit. In contrast, sonographers who had good scan completeness and image quality at baseline performed similarly in the follow‐up audit.
Reduction in variation in practice and sustaining good practice are both recognized factors in upholding patient safety10, 11; this has also been demonstrated in fetal ultrasound. Although not evaluated as part of this study, the observed individual improvements could enhance overall screening performance of an imaging department. It is important to note that the finding of improvement in scan completeness was observed not only overall but in each individual anatomical view, and that the overall scan completeness is lower than view completeness for each view, due to the rigorous way in which we defined scan completeness.
Interobserver agreement was high for the assessment of scan completeness; it was lowest for the head TV view, which may be because the original local protocol was unclear as to whether the head TV or transthalamic (TT) view should be stored, and therefore some recorded head views were in the TT view. Although TV and TT planes are very close to each other12, this could introduce some disagreement between reviewers. The interobserver agreement was higher for completeness compared with assessment of image quality. This is likely to be due to the more complex nature of image‐quality assessment, which includes assessing several criteria per view. It is also evident that assessment of caliper placement is not as reproducible as other criteria (Table 4). For instance, the agreement on the placement of AC calipers was less than 70% and the worst among other abdominal criteria.
One of the key strengths of this study is that the entire scan output of a single department was assessed, removing any potential biases in selection of examinations for review and offering the most comprehensive assessment of both departmental and individual performance. A limitation of this approach is that it is very labor intensive. Although it is possible that a retrospective audit of a small number (e.g. 10–20 cases) of consecutive scans would suffice, this would provide only limited understanding of individual and departmental performance. Another option, self‐scoring, in which sonographers appraise their own images, has been shown previously to be feasible13. However, this self‐scoring was undertaken within a framework of additional, independent scoring; it is not known whether this would be of benefit in the absence of such external review13.
In our study, independent visual assessment was performed by experienced sonographers. This was based on an objective, criteria‐based scoring system that was shown to have a high degree of reproducibility for image scoring. High reproducibility (kappa, 92.7) for image scoring methods shown in a previous study8 was also seen in the current study, and confirms the high level of training of the reviewers conducting the audit. Despite this, the assessment of placement of calipers in the four biometry views typically had a low agreement between reviewers (Table 4); the significant contribution of caliper placement to variability is in keeping with the findings of previous studies, such as that of Sarris et al.14. In addition, although the femur is a long straight object, there was significant disagreement between reviewers when assessing magnification of the femur images. In contrast, criteria that assessed if an object is not visible on an image, for example, no cerebellum in the head TV view, had high agreement between reviewers.
One potential limitation of this study was that the two audit cycles were back‐to‐back. This was to allow seamless observation of practice over time. It is highly likely that, had we allowed a period of time to pass before re‐audit to ensure that changes in practice in response to Cycle 1 were more embedded, greater improvements would have been seen. Given the described audit process, it is possible that some of the observed improvements in performance may have related more to sonographer awareness of ongoing audit and knowledge of the specific criteria used to evaluate performance. However, a longer study period could have been subject to confounding due to other unmeasured changes. Another limitation is that we assessed only six views, based on the recommendations for practice in the UK; we, of course, recognize that other scanning protocols suggest a larger number of images to be stored. However, the improvements seen in this smaller number of views should also apply. Although imaging of cardiac views was introduced in 2014, this was on the basis of best practice and they were not included in the analysis as they were not assessed in Cycle 1.
In conclusion, we have shown that large‐scale clinical audit, coupled with implementation of targeted changes and feedback to sonographers, can lead to improvements in image quality on the mid‐trimester anomaly scan, in terms of both completeness of scans and image quality. However, while quality improvement is possible, such comprehensive manual audit in a high‐throughput clinical setting is a very labor‐intensive process and this would be a major barrier for implementation in routine practice. Ongoing work on automated image analysis15 and further research into automated image recognition would open up the possibility of more rapid audit processes, and the current work provides evidence that this could be effective.
ACKNOWLEDGMENTS
This research was conducted as part of the AQABUS project, a joint partnership between the University of Oxford and Intelligent Ultrasound Ltd funded by Innovate UK (Project 101684) and the UK Engineering and Physical Sciences Research Council (EP/L505316/1). The manual audit tool was developed by Intelligent Ultrasound Ltd. J.A.N. and A.T.P. are cofounders of, and consult for, Intelligent Ultrasound Ltd, a company that develops software and systems to improve ultrasound imaging and services. A.T.P. is supported by the National Institute for Health Research (NIHR) Oxford Biomedical Research Centre (BRC). The support of the 12 reviewers who performed the audit is gratefully acknowledged.
REFERENCES
- 1. National Health Service (NHS) . Fetal Anomaly Screening Programme , 2018. https://assets.publishing.service.gov.uk/government/uploads/system/uploads/attachment_data/file/749742/NHS_fetal_anomaly_screening_programme_handbook_FINAL1.2_18.10.18.pdf.
- 2. Salomon LJ, Alfirevic Z, Berghella V, Bilardo C, Hernandez‐Andrade E, Johnsen SL, Kalache K, Leung K‐Y, Malinger G, Munoz H, Prefumo F, Toi A, Lee W. Practice guidelines for performance of the routine mid‐trimester fetal ultrasound scan. Ultrasound Obstet Gynecol 2011; 37: 116–126. [DOI] [PubMed] [Google Scholar]
- 3. Mahgerefteh S, Kruskal JB, Yam CS, Blachar A, Sosna J. Peer review in diagnostic radiology: current state and a vision for the future. Radiographics 2009; 29: 1221–1231. [DOI] [PubMed] [Google Scholar]
- 4. Royal College of Radiologists; Society and College of Radiographers. Standards for the provision of an ultrasound service, 2014. http://www.sor.org/sites/default/files/document‐versions/bfcr1417_standards_ultrasound.pdf. [Google Scholar]
- 5. Society and College of Radiographers; British Medical Ultrasound. Guidelines for professional ultrasound practice, 2015. https://www.sor.org/sites/default/files/document‐versions/ultrasound_guidance.pdf. [Google Scholar]
- 6. Cascade PN. Quality improvement in diagnostic radiology. Am J Roentgenol 1990; 154: 1117–1120. [DOI] [PubMed] [Google Scholar]
- 7. Kaewlai R, Abujudeh H. Peer review in clinical radiology practice. Am J Roentgenol 2012; 199: W158–W162. [DOI] [PubMed] [Google Scholar]
- 8. Salomon LJ, Bernard JP, Duyme M, Doris B, Mas N, Ville Y. Feasibility and reproducibility of an image‐scoring method for quality control of fetal biometry in the second trimester. Ultrasound Obstet Gynecol 2006; 27: 34–40. [DOI] [PubMed] [Google Scholar]
- 9. Salomon LJ, Winer N, Bernard JP, Ville Y. A score‐based method for quality control of fetal images at routine second‐trimester ultrasound examination. Prenat Diagn 2008; 28: 822–827. [DOI] [PubMed] [Google Scholar]
- 10. Chalouhi GE, Salomon LJ, Fontanges M, Althuser M, Haddad G, Scemama O, Chabot J‐M, Duyme M, Fries N. Formative assessment based on an audit and feedback improves nuchal translucency ultrasound image quality. J Ultrasound Med 2013; 32: 1601–1605. [DOI] [PubMed] [Google Scholar]
- 11. Jaudi S, Granger B, Herpin CN, Fries N, Du Montcel ST, Dommergues M. Online audit and feedback improve fetal second‐trimester four‐chamber view images: a randomised controlled trial. Prenat Diagn 2013; 33: 959–964. [DOI] [PubMed] [Google Scholar]
- 12. Napolitano R, Donadono V, Ohuma EO, Knight CL, Wanyonyi SZ, Kemp B, Norris T, Papageorghiou AT. Scientific basis for standardization of fetal head measurements by ultrasound: a reproducibility study. Ultrasound Obstet Gynecol 2016; 48: 80–85. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Cavallaro A, Ash ST, Napolitano R, Wanyonyi S, Ohuma EO, Molloholli M, Sande J, Sarris I, Ioannou C, Norris T, Donadono V, Carvalho M, Purwar M, Barros FC, Jaffer YA, Bertino E, Pang R, Gravett MG, Salomon LJ, Noble JA, Altman DG, Papageorghiou AT. Quality control of ultrasound for fetal biometry: results from the INTERGROWTH‐21st Project. Ultrasound Obstet Gynecol 2018; 52: 332–339. [DOI] [PubMed] [Google Scholar]
- 14. Sarris I, Ioannou C, Chamberlain P, Ohuma E, Roseman F, Hoch L, Altman DG, Papageorghiou AT. Intra‐ and interobserver variability in fetal ultrasound measurements. Ultrasound Obstet Gynecol 2012; 39: 266–273. [DOI] [PubMed] [Google Scholar]
- 15. Yaqub M, Kelly B, Papageorghiou AT, Noble JA. A deep learning solution for automatic fetal neurosonographic diagnostic plane verification using clinical standard constraints. Ultrasound Med Biol 2017; 43: 2925–2933. [DOI] [PubMed] [Google Scholar]