Table 3.
Number of BI (B.a. and G.s.) that tested positive after placement of SLE and fan
| Positive Biologic Indicators (+/total) | |||||||||||||||
| G. stearothermophilus | B. atrophaeus | ||||||||||||||
| BC | MR | RR | IIC | ICL | Total | BC | MR | RR | IIC | ICL | Total | ||||
| 43-L tote, 2.0 ga | ppm-h | ||||||||||||||
| No fan, no SLEb | 304 ± 127 | 2/25 | 1/25 | ||||||||||||
| SLE only | 120 ± 35 | 2 | 0 | 0 | 0 | 2 | 4/15 | 2 | 0 | 0 | 0 | 1 | 3/15 | ||
| SLE + Fanc | 18 ± 22 | 3 | 3 | 3 | 3 | 3 | 15/15 | 3 | 3 | 2 | 3 | 3 | 14/15 | ||
| Total | 5/6 | 3/6 | 3/6 | 3/6 | 5/6 | 19/30 | 5/6 | 3/6 | 2/6 | 3/6 | 4/6 | 17/30 | |||
| 43 L, 4.0 ga | ppm-h | ||||||||||||||
| No Fan, No SLEb | 923 ± 173 | 2/25 | 0/25 | ||||||||||||
| SLE | 442 ± 37 | 1 | 0 | 0 | 0 | 1 | 2/15 | 1 | 0 | 0 | 0 | 0 | 1/15 | ||
| SLE + Fanc | 115 ± 73 | 2 | 0 | 0 | 0 | 1 | 3/15 | 1 | 0 | 1 | 0 | 0 | 2/15 | ||
| Total | 3/6 | 0/6 | 0/6 | 0/6 | 2/6 | 5/30 | 3/6 | 0/6 | 0/6 | 0/6 | 0/6 | 3/30 | |||
| 75 L, 8.0 ga | ppm-h | ||||||||||||||
| SLE | 1225 ± 102 | 0 | 0 | 0 | 0 | 0 | 0/15 | 0 | 0 | 0 | 0 | 0 | 0/15 | ||
| Total | 0/0 | 0/0 | 0/0 | 0/0 | 0/0 | 0/15 | 0/0 | 0/0 | 0/0 | 0/0 | 0/0 | 0/15 | |||
Total product weight (g)
BI used but not associated with SLE, data is represented in Table 1
Small battery-powered fan
SLE, small lab equipment; BC, bottom of the calculator; MR, inside mouse restrainer; RR, inside rat restrainer; IIC, inside induction chamber; ICL, outside of induction chamber lid