. 2019 Sep;58(5):569–576. doi: 10.30802/AALAS-JAALAS-18-000105

Table 3.

Number of BI (B.a. and G.s.) that tested positive after placement of SLE and fan

Positive Biologic Indicators (+/total)
G. stearothermophilus B. atrophaeus
BC MR RR IIC ICL Total BC MR RR IIC ICL Total
43-L tote, 2.0 ga ppm-h
No fan, no SLEb 304 ± 127 2/25 1/25
SLE only 120 ± 35 2 0 0 0 2 4/15 2 0 0 0 1 3/15
SLE + Fanc 18 ± 22 3 3 3 3 3 15/15 3 3 2 3 3 14/15
Total 5/6 3/6 3/6 3/6 5/6 19/30 5/6 3/6 2/6 3/6 4/6 17/30
43 L, 4.0 ga ppm-h
No Fan, No SLEb 923 ± 173 2/25 0/25
SLE 442 ± 37 1 0 0 0 1 2/15 1 0 0 0 0 1/15
SLE + Fanc 115 ± 73 2 0 0 0 1 3/15 1 0 1 0 0 2/15
Total 3/6 0/6 0/6 0/6 2/6 5/30 3/6 0/6 0/6 0/6 0/6 3/30
75 L, 8.0 ga ppm-h
SLE 1225 ± 102 0 0 0 0 0 0/15 0 0 0 0 0 0/15
Total 0/0 0/0 0/0 0/0 0/0 0/15 0/0 0/0 0/0 0/0 0/0 0/15
a

Total product weight (g)

b

BI used but not associated with SLE, data is represented in Table 1

c

Small battery-powered fan

SLE, small lab equipment; BC, bottom of the calculator; MR, inside mouse restrainer; RR, inside rat restrainer; IIC, inside induction chamber; ICL, outside of induction chamber lid