Fig. 4.
Analytical performance of ND-COOH 15/CFME for dopamine detection. (A) log-log relationship between anodic peak current of 1 μM dopamine and scan rate (n = 3). (B) Concentration dependence from the anodic peak current of 20 nM to 100 μM dopamine in PBS pH 7.4 (n = 4), compared to unmodified CFME. (C) Linear range at both electrodes. (D) Example CV of 20 nM dopamine in PBS pH 7.4 with current-time trace inset for 5-s bolus injection.