Table 1.
Sample | Spiked [MC-LR] μg/L | ELISA μg/L | SERS immunoassay μg/L | RSD (%) | Recovery percentage |
---|---|---|---|---|---|
Lake+sample 0 | 0 | undetectable | 0.03 | 17 | NA |
Lake+sample 1 | 0.21 | 0.23 | 0.22 | 14 | 105% |
Lake+sample 2 | 0.76 | 0.73 | 0.81 | 8.0 | 107% |
Lake+sample 3 | 6.10 | 6.40 | 6.12 | 1.8 | 100% |
Note: Lake+sample 0 is the Mystic Lake surface water collected from the Upper Mystic Lake in Massachusetts without the addition of MC-LR; Lake+samples 1, 2 and 3 were generated by spiking with the stock solution of MC-LR into the Mystic Lake water sample; RSD is the relative standard deviation.