Table 2. Intra-day reproducibility for BPA in human plasma.

Day Quality Control Average SD % CV Accuracy (%) Inaccuracy (%)
1 QCL (70 ng/mL) 74.68 2.45 3.28 106.6 6.6
QCM (800 ng/mL) 755.8 21.84 2.89 94.48 5.52
QCH (1500 ng/mL) 1426 105.26 7.38 95 5
2 QCL (70 ng/mL) 61.4 5.05 8.22 87.72 12.28
QCM (800 ng/mL) 723 32.53 4.5 90.38 9.62
QCH (1500 ng/mL) 1434 96.7 6.7 95.68 4.32
3 QCL (70 ng/mL) 62.82 4.69 7.47 89.76 10.24
QCM (800 ng/mL) 770.2 72.78 9.45 96.32 3.68
QCH (1500 ng/mL) 1576 134.65 8.54 105.08 5.08

QCL: quality control low; QCM: quality control medium; QCH: quality control high; SD: standard deviation; CV: coefficient of variation