Table 2. Intra-day reproducibility for BPA in human plasma.
Day | Quality Control | Average | SD | % CV | Accuracy (%) | Inaccuracy (%) |
---|---|---|---|---|---|---|
1 | QCL (70 ng/mL) | 74.68 | 2.45 | 3.28 | 106.6 | 6.6 |
QCM (800 ng/mL) | 755.8 | 21.84 | 2.89 | 94.48 | 5.52 | |
QCH (1500 ng/mL) | 1426 | 105.26 | 7.38 | 95 | 5 | |
2 | QCL (70 ng/mL) | 61.4 | 5.05 | 8.22 | 87.72 | 12.28 |
QCM (800 ng/mL) | 723 | 32.53 | 4.5 | 90.38 | 9.62 | |
QCH (1500 ng/mL) | 1434 | 96.7 | 6.7 | 95.68 | 4.32 | |
3 | QCL (70 ng/mL) | 62.82 | 4.69 | 7.47 | 89.76 | 10.24 |
QCM (800 ng/mL) | 770.2 | 72.78 | 9.45 | 96.32 | 3.68 | |
QCH (1500 ng/mL) | 1576 | 134.65 | 8.54 | 105.08 | 5.08 |
QCL: quality control low; QCM: quality control medium; QCH: quality control high; SD: standard deviation; CV: coefficient of variation