Table 5. Autosampler stability of BPA in human plasma.

Stability Quality Control Average SD % CV Accuracy (%) Inaccuracy (%)
After 24 h QCL (70 ng/mL) 64.02 2.5 3.9 91.46 8.54
QCM (800 ng/mL) 745 37.30 5.01 93.14 6.86
QCH (1500 ng/mL) 1500 75.83 5.06 100.08 0.08

QCL: quality control low; QCM: quality control medium; QCH: quality control high; SD: standard deviation; CV: coefficient of variation