23 |
Low yield of TFA-NHP |
Degraded or contaminated reagents used
during synthesis. |
Use fresh, high purity reagents.
Trifluoroacetic acid anhydride is prone to hydrolysis and should be used
within a week of opening. |
42 |
Low yield of cross-linker during resin
cleavage, resin has uncleaved cross-linker remaining. |
Residual DMF on cross-linker resin inhibits
cleavage. |
Wash resin extensively with DCM, and then
incubate the resin in DCM for 20 minutes, pipetting occasionally to
improve solvent accessibility. Make a fresh stock of cleavage solution
and repeat the incubation in cleavage solution. |
42 |
Impure cross-linker observed during direct
infusion ESI-MS analysis. |
Degraded or contaminated reagents used
during synthesis. |
Make sure all solvents and reagents used
for synthesis are freshly made and of high purity. Solvents should be
ACS grade or higher. DMF should not be more than 6 months old. Amino
acid and DIC solutions are stable for up to two weeks. The piperidine
solution is stable for one month. |
44 |
Cross-linker has hydrolyzed. |
Cross-linker stock solution has become
contaminated with water. |
Verify hydrolysis by ESI-MS anlaysis. See
Fig. 4. |
65 |
Cross-linker precipitated in pipet
tip. |
Pipet tip was immersed into aqueous buffer
prior to expelling the concentrated cross-linker solution causing it to
precipitate |
Do not immerse pipet tip in the solution
when adding cross-linker. Instead hold the tip above the solution and
proceed with pushing the cross-linker solution out. |
66 |
No yellow color observed during
cross-linking reaction. |
pH of solution is too low. |
Check the pH of the reaction solution. It
should be between 7–8. If it is lower adjust it within range by
titrating with 1M NaOH or use a higher concentration of buffer. |
73 |
Cell pellet is small or solution is sticky
and viscous. |
Cells have lysed during pelleting and
resuspension. |
Check centrifugation speed to make sure it
is appropriate for the sample type used. Be extremely gental while using
a pipet to resuspend cells only using a minimal number of cycles to
suspend the cell pellet. |
75 |
Tissue samples not lysing during
cryogrinding. |
Cryoginding ball was added to liquid sample
prior to freezing in liquid nitrogen and is now encased in ice and not
able to grind sample. |
Add cryogriding ball after sample solution
has been frozen in liquid N2 to ensure it is free to move during
shaking. |
80 |
Large amount of precipitation in tryptic
digest sample. |
Typsin failed to digest proteins into
peptides. |
Sample solution conditions were not
compatible with trypsin. Check pH to ensure it is between 7–8.
Ensure urea concentration is less than 1 M. Correct solution conditions
and add fresh trypsin to repeat step 80. |
81 |
Large amount of precipitation observed
after acidifying sample. |
Large amount of residual undigested protein
due to inefficient trypsin digestion |
Check solution conditions for trypsin
digestion. Make sure the pH is between 7–8, urea concentration is
less than 1 M and fresh trypsin is used. Repeat steps 80–81. |
97 |
No signal or low signal observed in SCX
absorbance traces. |
Loss of protein during desalting due to
poor binding to the packing material. |
Check pH of the input sample for the
Sep-Pak. Reacidify if above 3, and repeat the desalting procedure. |
111 |
Spray is sputtering or droplets forming at
the tip of the column. |
Incorrect spray voltage or tip position,
blunt, broken or dirty spray tip. |
Adjust the applied spray voltage and tip
position until stable spray is obtained. If tip appears blunt or dirty
replace with a new column. |
112 |
No signal or low signal
observed in chromatogram on mass spectrometer. |
Poor binding of sample to monomeric avidin
resin. |
Verify that the pH of the SCX fractions
after base addition are approximately 8. These samples can be recaptured
again after the pH is corrected. |
Sample not injected by autosampler on
LC |
Verify that autosampler is delivering the
requeseted volume of sample onto the LC column. |
Mass spectrometer is out of calibration or
has low sensitivity due to dirty ion optics. |
Clean and calibrate mass spectrometer
verifying sensitivity with a standard sample. |
Multiple contributing factors. |
Use a cross-linked purified protein such as
BSA as a positive control sample and ensure that results similar to
those presented Fig. 5 and Table 2 are obtained. |
120 |
No significant identifications in .pep.xml
file. |
Incorrect database or search
parameters. |
Check that the correct protein database was
used and the the comet search parameters match the experimental
setup. |