. 2019 Oct 4;5(10):eaay1394. doi: 10.1126/sciadv.aay1394

Table 1. Resonator EPR signal characteristics calculated and measured using a power saturation measurement of a lithium phthalocyanine point sample.

Dimensions of the resonators and further experimental details are provided in the Supplementary Materials.

Geometry Unsat.
signal
Sat.
signal
Eff.
η
Q0-value
Meas.
ηQ0
Calc. Meas. Calc. Meas.
Bruker MD5W1 1.0 1.0 1.0 1.0 6.12 × 10−6 6650 1.0
Bruker MS3 1.5 1.2 1.0 1.0 79.6 × 10−6 600 1.17
PMR RO6010LM 4.4 1.2 0.9 1.2 4.63 × 10−3 61 6.94
PMR sapphire 18.6 13.3 3.9 3.8 4.63 × 10−3 181 20.59
Microhelix 35.7 28.2 6.1 5.7 6.63 × 10−3 220 35.84