. 2019 Jul 10;9(3):71. doi: 10.3390/diagnostics9030071

Table 2.

Intra-rater and inter-rater reliability of craniocervical ranges measured by the hybrid device.

Intra-Rater ICC (95% CI) Intra-Rater SEM (Degrees) Intra-Rater MDC (Degrees) Inter-Rater ICC (95% CI) Inter-Rater SEM (Degrees) Inter-Rater MDC (Degrees)
Flexion 0.963
(0.933 to 0.980)
2.68 7.42 0.979
(0.962 to 0.989)
1.81 5.01
Extension 0.954
(0.916 to 0.975)
2.50 6.92 0.953
(0.914 to 0.975)
2.61 7.24
Side-bending to right 0.925
(0.864 to 0.959)
1.60 4.44 0.880
(0.786 to 0.934)
2.03 5.63
Side-bending to left 0.939
(0.888 to 0.967)
1.48 4.12 0.931
(0.875 to 0.963)
1.59 4.39
Rotating to right 0.959
(0.925 to 0.978)
1.66 4.61 0.942
(0.895 to 0.969)
1.96 5.44
Rotating to left 0.931
(0.874 to 0.962)
1.99 5.52 0.956
(0.919 to 0.976)
1.63 4.51

Note: CI, confidence interval; ICC, intraclass correlation coefficient (analyzed by the two-way random effect model); SEM, standard error of measurement; MDC, minimal detectable change.