Table 3. Postoperative arithmetic and absolute refractive errors of the two SS-OCT devices (n = 14) in eyes which AXL was not measured by the PCI device.
IOLMaster 700 | Argos | p value | |
---|---|---|---|
Mean arithmetic prediction error (D) |
0.10 ± 0.45 | 0.15 ± 0.37 | 0.453 a |
Mean absolute prediction error (D) |
0.33 ± 0.32 | 0.34 ± 0.22 | 0.984 a |
Median absolute predication error (D) | 0.22 | 0.28 | - |
Eyes within ± 0.5 D (%) | 78.57 | 71.43 | 0.668 b |
a Paired t-test