Skip to main content
. 2019 Aug 28;2(3):75. doi: 10.3390/mps2030075

Table 1.

Characteristics of the described 241Am source and current setup. Adapted from [24]. Used for calculating the irradiation time.

Characteristic Value
Area of active surface (mm2) 11
Source fluence at cell position (particles/s /cm2) 15966
Half-life (days) 157800
Distance between the active surface and mylar (mm) 5.0 (± 0.1)
Distance between collimator and Culture dish holder (mm) 2.8 (± 0.1)
Irradiation time per point (s/Gy) 1 240.1 (± 5.9%)

1 Calculated in an 8 µm layer of water above a 1.4 µm Mylar sheet, irradiated in 2017 using the described 241Am source.