Table 2. Diagnostic Performance of S-SWE and TE in Differentiation of Histological Stages of Hepatic Fibrosis in 107 Patients.
Histological Results | Diagnostic Performance | S-SWE | TE | P* |
---|---|---|---|---|
≤ F1 vs. ≥ F2 | AUC (95% CI) | 0.842 (0.754–0.930) | 0.850 (0.768–0.931) | 0.788 |
Optimism-corrected AUC | 0.844 | 0.851 | N.A. | |
≤ F2 vs. ≥ F3 | AUC (95% CI) | 0.844 (0.771–0.916) | 0.841 (0.767–0.915) | 0.914 |
Optimism-corrected AUC | 0.845 | 0.843 | N.A | |
≤ F3 vs. F4 | AUC (95% CI) | 0.850 (0.775–0.925) | 0.868 (0.800–0.936) | 0.475 |
Optimism-corrected AUC | 0.85 | 0.867 | N.A. |
*p values were calculated using Delong test. S-SWE; Samsung Medison Co., Ltd. AUC = area under receiver operating characteristic curve, CI = confidence interval, N.A. = not applicable, S-SWE = S-shearwave elastography