Skip to main content
. 2019 Oct 22;7(11):E1327–E1332. doi: 10.1055/a-0957-3005

Table 4. Comparison of diagnostic accuracy between Franseen needle and conventional needle groups.

Propensity-matched patients
Diagnostic accuracy, n (%) Group C Group F P
Location
  • Pancreatic head

19/24 (79.2 %) 48/49 (98.0 %) 0.013
  • Body or tail

63/71 (88.7 %) 39/46 (84.8 %) 0.58
Puncture route
  • Transgastric

64/72 (88.9 %) 41/48 (85.4 %) 0.59
  • Transduodenal

18/23 (78.3 %) 46/47 (97.9 %) 0.01
Tumor type
  • Adenocarcinoma

64/77 (83.1 %) 57/64 (89.1 %) 0.34
  • Special pancreatic tumors

10/10 (100 %) 16/17 (94.1 %) > 0.99
  • Benign lesion

8/8 (100 %) 14/14 (100 %) > 0.99
Tumor size
  • ≥ 20 mm

62/72 (86.1 %) 71/76 (93.4 %) 0.18
  • < 20 mm

20/23 (87.0 %) 16/19 (84.2 %) > 0.99
  • Overall

82/95 (86.3 %) 87/95 (91.6 %) 0.36

n, number of lesions