Table 1.

Characteristics of the Discovery MI and Gemini TF PET-CT systems

PET-CT system Discovery MI Gemini TF
Crystal material LYSO LYSO
Amplifier SiPM PM-tubes
Number of rings 4 3
Size of crystals [mm3] 3.95 × 5.3 × 25 4.0 × 4.0 × 22
Axial FOV [mm] 200 180
Bore diameter [mm] 700 700
Overlap [%] 23.9 50
NEMA Sensitivity [cps/kBq] 13.8 7.0
FWHM axial @1 cm [mm] 4.7 5.8
Coincidence window width (ns) 4.9 6
Timing resolution (ps) 382 600
Lower energy threshold (keV) 425 440
Matrix 256 × 256 144 × 144
Pixel size (mm2) 2.7 × 2.7 4.0 × 4.0
Slice thickness [mm] 2.79 4.0
Image planes in the axial FOV 71 45
Number of CT slice 128 16

FOV field of view; FWHM full width at half maximum; LYSO lutetium-yttrium oxyorthosilicate; NEMA National Electrical Manufacturers Association; PM photomultiplier; SiPM silicon photomultiplier