. 2019 Oct 8;20(10):172–180. doi: 10.1002/acm2.12734

Table 2.

Fluoroscope loop technique and Shewhart Control Chart Data from all five IR rooms using the CD phantom.

  IR‐15 IR‐16 IR‐17 IR‐18 IR‐19
N days 160 157 154 158 158
Mean kVp 65.7 66.0 67.1 65.9 71.3
kVp SD 1.0 0 1.5 0.28 0
Mean mA 227.3 144.9 175.2 205.9 97.9
mA SD 43.2 1.1 58.3 6.4 0.4
Mean SNR 49.8 50.9 48.1 23.6 61.2
SD SNR 1.21 0.66 1.08 0.92 1.05
CV SNR 2.4% 1.3% 2.2% 3.9% 1.7%
# instances above UCL 1 0 0 0 0
# instances below LCL 0 1 0 3 1

Abbreviations: SNR, signal to noise ratio; CNR, contrast to noise ratio; CV, coefficient of variation; UCL, upper control limit.