Table 1.
The Results of K, Na, and Cl for SRM 956c in Six Selected Labs of China
956c‐I | 956c‐II | 956c‐III | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Electrolyte | Lab | Target | Measurement | Bias (%) | Target | Measurement | Bias (%) | Target | Measurement | Bias (%) | Mean bias (%) | Standarda (%) |
K | 1 | 5.976 | b | b | 3.977 | 3.89 | −2.19 | 1.982 | 1.90 | −4.14 | −3.17c | ±2 |
(mmol/l) | 2 | 5.83 | −2.44 | 3.95 | −0.68 | 2.22 | 12.01 | 2.96c | ||||
3 | 6 | 0.40 | 4.1 | 3.09 | 2.10 | 5.95 | 3.15c | |||||
4 | 6.04 | 1.07 | 3.98 | 0.08 | 2.06 | 3.94 | 1.70 | |||||
5 | 6.03 | 0.90 | 4.02 | 1.08 | 2.04 | 2.93 | 1.64 | |||||
6 | 6.06 | 1.41 | 4.04 | 1.58 | 2.02 | 1.92 | 1.64 | |||||
Na | 1 | 118.8 | b | b | 137.5 | 137.0 | −0.33 | 157.4 | 152.2 | −3.29 | −1.81c | ±1.5 |
(mmol/l) | 2 | 120.3 | 1.23 | 136.9 | −0.41 | 161.5 | 2.63 | 1.15 | ||||
3 | 121.6 | 2.36 | 139.0 | 1.09 | 154.4 | −1.91 | 0.51 | |||||
4 | 120.2 | 1.18 | 139.0 | 1.09 | 166.2 | 5.59 | 2.62c | |||||
5 | 120.7 | 1.57 | 139.3 | 1.29 | 158.0 | 0.36 | 1.07 | |||||
6 | 123.1 | 3.60 | 140.5 | 2.15 | 154.2 | −2.05 | 1.23 | |||||
Cl | 1 | 104.9 | b | b | 121.5 | 119.0 | −2.09 | 137.4 | 133.8 | −2.65 | −2.37c | ±1.5 |
(mmol/l) | 2 | 102.9 | −1.89 | 118.0 | −2.88 | 140.6 | 2.34 | −0.81 | ||||
3 | 103.4 | −1.43 | 120.0 | −1.23 | 132.0 | −3.93 | −2.20c | |||||
4 | 103.3 | −1.56 | 119.4 | −1.71 | 143.4 | 4.38 | 0.37 | |||||
5 | 102.4 | −2.35 | 119.2 | −1.93 | 135.8 | −1.19 | −1.82 | |||||
6 | 104.3 | −0.59 | 120.8 | −0.58 | 134.3 | −2.27 | −1.15 |
Health industry standards of China 7 were applied as the standard.
Due to rupture of the container, Lab 1 could not provide the data.
The bias is beyond the range of the standard.