. 2014 May 5;29(3):226–234. doi: 10.1002/jcla.21755

Table 1.

The Results of K, Na, and Cl for SRM 956c in Six Selected Labs of China

956c‐I 956c‐II 956c‐III
Electrolyte Lab Target Measurement Bias (%) Target Measurement Bias (%) Target Measurement Bias (%) Mean bias (%) Standarda (%)
K 1 5.976 b b 3.977 3.89 −2.19 1.982 1.90 −4.14 −3.17c ±2
(mmol/l) 2 5.83 −2.44 3.95 −0.68 2.22 12.01 2.96c
3 6 0.40 4.1 3.09 2.10 5.95 3.15c
4 6.04 1.07 3.98 0.08 2.06 3.94 1.70
5 6.03 0.90 4.02 1.08 2.04 2.93 1.64
6 6.06 1.41 4.04 1.58 2.02 1.92 1.64
Na 1 118.8 b b 137.5 137.0 −0.33 157.4 152.2 −3.29 −1.81c ±1.5
(mmol/l) 2 120.3 1.23 136.9 −0.41 161.5 2.63 1.15
3 121.6 2.36 139.0 1.09 154.4 −1.91 0.51
4 120.2 1.18 139.0 1.09 166.2 5.59 2.62c
5 120.7 1.57 139.3 1.29 158.0 0.36 1.07
6 123.1 3.60 140.5 2.15 154.2 −2.05 1.23
Cl 1 104.9 b b 121.5 119.0 −2.09 137.4 133.8 −2.65 −2.37c ±1.5
(mmol/l) 2 102.9 −1.89 118.0 −2.88 140.6 2.34 −0.81
3 103.4 −1.43 120.0 −1.23 132.0 −3.93 −2.20c
4 103.3 −1.56 119.4 −1.71 143.4 4.38 0.37
5 102.4 −2.35 119.2 −1.93 135.8 −1.19 −1.82
6 104.3 −0.59 120.8 −0.58 134.3 −2.27 −1.15
a

Health industry standards of China 7 were applied as the standard.

b

Due to rupture of the container, Lab 1 could not provide the data.

c

The bias is beyond the range of the standard.