Table 2.

The Total CV (%) of K, Na, and Cl for Roche Quality Control Materials in Six Selected Labs

Analyte Electrolyte Lab 1 Lab 2 Lab 3 Lab 4 Lab 5 Lab 6 Opt Des Min Standarda
Precipath U K 0.99 1.36 1.43 1.79 0.82 b 1.2 2.4 3.6 2.5
Na 0.96 1.24 1.16 1.05 0.66 b 0.2 0.4 0.5 1.5
Cl 1.85a 1.17 1.53a 1.18 1.65a b 0.3 0.6 0.9 1.5
Precinorm U K 1.26 1.97 1.57 1.91 0.9 1.31 1.2 2.4 3.6 2.5
Na 1.09 1.49 1.41 1.47 0.6 1.15 0.2 0.4 0.5 1.5
Cl 1.96a 1.54a 2.09a 1.66a 1.79a 1.36 0.3 0.6 0.9 1.5
a

Health industry standards of China 7 were applied as the standard.

b

Due to misoperation of the test, Lab 6 could not provide the data.

Opt, optimal requirement of group biological variation 6; Des, desirable requirement of group biological variation 6; Min, minimal requirement of group biological variation 6.