Skip to main content
. 2015 Aug 24;30(5):424–430. doi: 10.1002/jcla.21874

Table 2.

Diagnostic Accuracy of Erythrocyte and Leukocyte Detection Using URiSCAN Devices and Roche Urine Analyzers With Manual Microscopy as the Standard

Erythrocytes Leukocytes
Sensitivity Specificity AUC P Sensitivity Specificity AUC P
URiSCAN Optima 91.4 78.1 0.848 0.931 76.6 93.1 0.790 0.003a
Urisys 1100 86.7 82.6 0.846 64.9 97.8 0.740
URiSCAN Pro II 79.6 85.8 0.827 0.400 72.1 93.7 0.795 0.746
Cobas u411 76.8 86.4 0.816 70.9 94.6 0.790
URiSCAN Super plus 92.9 83.8 0.883 0.018a 87.5 84.8 0.850 0.008a
Urisys 2400 92.9 81.1 0.870 95.8 61.5 0.770
a