Skip to main content
. 2016 Feb 3;30(5):663–671. doi: 10.1002/jcla.21919

Table 6.

Comparison of the Pairwise Results for WBC among the Two Instruments and Manual Microscopy

Manual microscopy (counts/HPF)
cobas u 701 (counts/HPF) 0–1 2–4 5–10 11–20 21–30 >30 Sum
0–1 67 25 6 0 0 0 98
2–4 4 12 20 7 0 1 44
5–10 1 4 19 12 1 3 40
11–20 2 0 5 10 10 6 33
21–30 0 0 0 2 1 3 6
>30 0 1 0 1 0 34 36
Manual microscopy (counts/HPF)
UF‐1000i (counts/HPF) 0–1 2–4 5–10 11–20 21–30 >30 Sum
0–1 68 15 0 0 0 0 83
2–4 3 21 18 0 0 0 42
5–10 3 4 26 16 1 0 50
11–20 0 1 2 14 8 6 31
21–30 0 0 2 2 3 8 15
>30 0 1 2 0 0 33 36
cobas u 701 (counts/HPF)
UF‐1000i (counts/HPF) 0–1 2–4 5–10 11–20 21–30 >30 Sum
0–1 73 8 1 1 0 0 73
2–4 18 20 4 0 0 0 18
5–10 6 11 23 10 0 0 6
11–20 0 4 7 17 3 0 0
21–30 0 0 5 4 3 3 0
>30 1 1 0 1 0 33 1

(Inline graphic) represents the number of cases within one grade of difference and (Inline graphic) the number with the same grade.