Skip to main content
. 2014 Mar 20;28(6):419–427. doi: 10.1002/jcla.21704

Table 1.

Imprecision Tests of FL‐ICA System for Cardiac Biomarkers

Intra‐assay Interassay
Sample (μg/l) Meana SD CV% Meana SD CV%
A. Myoglobin
10.00 9.74 0.66 6.80 9.75 0.67 6.93
50.00 49.64 2.60 5.24 49.84 2.82 5.65
100.00 102.04 5.44 5.33 101.76 5.81 5.71
190.00 192.40 8.08 4.19 193.41 10.40 5.37
325.00 323.52 12.25 3.79 325.98 14.34 4.39
B. CK‐MB
3.00 3.37 0.25 7.32 3.61 0.28 7.72
10.00 10.15 0.58 5.71 10.28 0.62 6.03
25.00 25.19 1.35 5.37 25.37 1.50 6.01
50.00 50.57 2.80 5.54 50.87 2.91 5.72
90.00 90.15 4.25 4.71 90.05 4.99 5.54
C. cTnI
0.53 0.54 0.03 6.36 0.51 0.04 7.64
1.50 1.51 0.10 6.33 1.54 0.11 7.48
3.57 3.57 0.22 6.16 3.58 0.24 6.70
11.50 11.51 0.49 4.26 11.48 0.53 4.62
45.00 45.09 1.78 4.00 45.39 1.85 4.09
a