Table 2.
Validation of rRT‐PCR Array Sensitivity
Ct value | ||||
---|---|---|---|---|
Assay | Number of RNA (copies/reaction tube) | RT‐PCR array | Individual RT‐PCR | Standard PCR |
HA | 100,000 | 21.79 | 21.19 | + |
10,000 | 25.54 | 26.07 | + | |
1,000 | 30.78 | 29.77 | + | |
100 | 34.16 | 33.84 | − | |
10 | 37.67 | 37.23 | − | |
1 | − | − | − | |
H1N1 | 100,000 | 19.50 | 19.77 | + |
10,000 | 22.50 | 22.61 | + | |
1,000 | 25.13 | 26.48 | + | |
100 | 28.34 | 28.30 | − | |
10 | 35.37 | 34.97 | − | |
1 | − | − | − | |
H5N1 | 100,000 | 21.77 | 21.46 | + |
10,000 | 25.33 | 25.43 | + | |
1,000 | 29.13 | 29.38 | + | |
100 | 33.54 | 32.73 | + | |
10 | 37.08 | 36.36 | − | |
1 | − | − | − | |
H9N2 | 100,000 | 19.97 | 20.14 | + |
10,000 | 23.28 | 23.49 | + | |
1,000 | 27.33 | 27.41 | + | |
100 | 31.24 | 31.38 | − | |
10 | 36.04 | 35.18 | − | |
1 | − | − | − |
Positive (+) and negative detections (−) are indicated