Skip to main content
. 2013 Jul 12;27(4):312–316. doi: 10.1002/jcla.21604

Table 3.

The coefficients of variation of the microscopic analysis (UriSed, Fus100 and manual microscopy)

Within precision Between precision
Level 1 Level 2 Level 1 Level 2
Method Cells Mean ± SD CV(%) Mean ± SD CV(%) Mean ± SD CV(%) Mean ± SD CV(%)
UriSed Erythrocyte 6,4 ± 0,5 7,8 61,1 ± 2,5 4,1 3,9 ± 1,1 28,2 22,6 ± 2,1 9,2
Leukocyte 7,6 ± 1,0 13,1 19,9 ± 2,9 14,5 2.8 ± 0,8 28.5 6,7 ± 1,0 14,9
Fus100 Erythrocyte 3.1 ± 0.6 19.3 25.9 ± 1.3 5.1 0.0 ±a a 97.0 ± 4.6 4.7
Leukocyte 3.3 ± 0.7 21.2 14.8 ± 1.5 10.1 0.0 ±a a 5.9 ± 1.1 18.6
Manual Erythrocyte 2.7 ± 0.9 33.3 15.4 ± 2.1 13.6 2.3 ± 0.7 30.4 14.2 ± 2.9 20.4
Leukocyte 4.2 ± 1.1 26.1 27.0 ± 2.3 8.5 2.0 ± 0.7 35.0 22.4 ± 2.8 12.5

CV, Coefficients of variations, SD, Standard deviation.

a

SD and CV did not get calculated because the mean value of erythrocyte and leukocyte results was 0.