Table 1.

Precision of the Freelite™ FLC Assay on the SPAPLUS™ Using Two Concentration Levels of Control Material (n = 20)

Test result Claim of manufacturer Allowable imprecisiona
Mean Within‐run imprecision Total imprecision Within‐run imprecision Total imprecision
Analyte (mg/l) SD 95% CI %CV SD 95% CI %CV SD %CV SD %CV SD %CV
κFLC 15.1 0.51 0.358–0.898 3.40 0.72 0.505–1.275 4.80 0.319 2.11 1.584 10.49 0.50 3.30
29.4 0.99 0.694–1.742 3.40 1.23 0.922–1.850 4.20 0.483 1.64 2.764 9.40 0.97
λFLC 30.0 0.40 0.279–0.701 1.30 1.33 0.866–2.823 4.40 0.754 2.51 1.495 4.98 1.58 5.25
60.1 1.13 0.791–1.988 1.90 2.96 1.872–6.945 4.90 1.356 2.26 2.941 4.89 3.16
a

Minimum specifications for imprecision derived from biologic variation of κFLC and λFLC.