Skip to main content
. 2014 Jan 29;28(3):229–236. doi: 10.1002/jcla.21671

Table 3.

Carryover and Drift From Multifactor Regression Showing Components of Measurement Error in Preliminary Evaluation of the Freelite™ FLC Assay on the SPAPLUS™

Analyte (mg/l) Components of measurement error Mean of five runs Sign statistic P a
κFLC Carryover 1.2% 5 0.0625
Drift −0.018 2 1.0000
λFLC Carryover 0.0% 2 1.0000
Drift −0.230 1 0.3750
a

The sign test was statistically significant at 6.25% significance level.