Table 1.

Acquisition parameters for the mpMRI of the prostate with endorectal coll.

Series PSD Scan plane TR (ms) TE (ms) Slice/gap (mm) FOV (mm) Acquisition matrix NEX Sequence specific
Scout FSE 3-plane 867 83 5/1.5 400 × 400 256 × 192 1
T1 FGRE Axial 5.06 2.46 4.2/0 240 × 240 192 × 128 1 3D
T2 FSE Oblique axial 5000 96 3/0 180 × 180 256 × 256 3 2D
T2 FSE/CUBE Oblique axial 2400 142.5 1.6/0 180 × 180 256 × 224 1 3D reformatting recommended
DWI mid ss-EPI Oblique axial 4725 Min 3/0 180 × 180 128 × 64 6 b = 600 s/mm2; rFOV recommended
DWI high ss-EPI Oblique axial 4725 Min 3/0 260 × 260 128 × 64 7 b = 1350 s/mm2
DCE 3D SPGR Oblique axial Min Min 3/0 260 × 260 192 × 128 1 Temporal resolution = 10 s