Table 3. Device Performance Metrics at Implantation and Follow up .

Device Performance metrics from interrogations from those in whom quantitative interrogations were available. BPM=beats per minute; mV=millivolts.

HD SD P-level
At Implantation (N=132) (N=22)
Final Pacing Impedance (Ohms) 699 (±196) 768 (±177) 0.127
Final Pacing Capture Threshold (Volts) 0.631 (±0.469) 0.685 (±0.434) 0.620
Pulse Width (milliseconds) 0.329 (±0.216) 0.317 (±0.109) 0.816
Final Sensing Amplitude (in mV) 10.753 (±5.408) 10.996 (±5.554) 0.846
Lower Rate (BPM) 62 (±12) 60 (±12) 0.515
At Follow Up (N=103) (N=22)
Follow up Pacing Impedance (Ohms) 614 (±123) 624 (±107) 0.729
Follow up Pacing Capture Threshold (Volts) 0.633 (±0.601) 0.768 (±0.495) 0.328
Pulse Width (milliseconds) 0.31 (±0.09) 0.3 (±0.08) 0.699
Follow up sensing amplitude (mV) 12.625 (±5.778) 13.532 (±7.12) 0.598
Follow Up Lower Rate 62 (±11) 60 (±10) 0.462
Changes from Implantation to Follow Up (N=98) (N=19)
Change in Impedance (Ohms) -98 (±171) -155 (±100) 0.166
Change in Capture Threshold (Volts) 0.014 (±0.570) 0.016 (±0.227) 0.992
Change in Pulse Width (milliseconds) -0.018 (±0.263) -0.015 (±0.127) 0.965
Change in Sensing Amplitude (mV) 2.28 (±4.88) 2.59 (±4.68) 0.802
Change in Impedance (%) -16.88 (±29.25) -23.63(±15.06) 0.331
Change in Capture Threshold (%) -17.78 (±73.69) -4.98 (±35.93) 0.255
Change in Pulse Width (%) -15.65 (±105.48) -10.49 (±44.98) 0.844
Change in Sensing Amplitude (%) 4.59 (±89.11) 2.56 (±69.07) 0.928