Table I.
Methoda | Directionb | Tumor Localization Errorc |
---|---|---|
FDK | x | 1.2 ± 0.4 (0.4, 2.0) |
y | 1.2 ± 0.4 (0.1, 2.3) | |
z | 1.5 ± 0.4 (0.7, 2.5) | |
d | 2.3 ± 0.5 (1.5,3.5) | |
TV | x | 1.1 ± 0.3 (0.4, 1.7) |
y | 1.2 ± 0.4 (0.5, 2.0) | |
z | 1.5 ± 0.4 (0.8, 3.1) | |
d | 2.3 ± 0.4 (1.5, 3.6) | |
TV+SR-CNN | x | 0.4 ± 0.3 (0.0, 1.0) |
y | 0.5 ± 0.3 (0.0, 1.0) | |
z | 0.7 ± 0.3 (0.1, 1.1) | |
d | 1.1 ± 0.2 (0.4, 1.4) |
FDK is the under-sampled image reconstructed using FDK with ramp filter, TV is the under-sampled image reconstructed using TV regularization, and SR-CNN is the augmented TV image using the proposed SR-CNN.
x is the left-right (LR) direction, y is the anterior-posterior (AP) direction, z is the superior-inferior (SI) direction, and d is the 3D vector length of the errors.
tumor shift unit is millimeter (mm). The results are presented in mean ± standard deviation of the absolute errors, and the minimum and maximum absolute errors are listed in the brackets.