Table 1.

Performance criteria for closed-loop vasopressor controller in stable sepsis simulation scenarios with differing norepinephrine pharmacodynamics

Norepinephrine response level MDPE (%) MDAPE (%) Wobble (%) Divergence (%) m−1 Oscillation (cycles h−1) Oscillation amplitude (mmHg)
0.1 × 1.0 1.2 0.5 − 7.1 - -
0.2 × 0.9 1 0.6 − 4.6 - -
0.5 × 0.7 0.8 0.4 − 2.3 - -
1 × 0.7 0.7 0.3 − 1.2 - -
2 × 0.6 0.7 0.4 − 1 - -
5 × 0.7 0.7 0.4 − 0.3 - -
10 × 1.2 1.2 1.0 − 0.1 18–22 4

MDPE median performance error, MDAPE median absolute performance error; definitions (including wobble and divergence) found in Varvel et al. [30]

Oscillation and Oscillation Amplitude—the frequency of cycling of the MAP around the target, and the total pressure change (maximum–minimum) during a cycle